Product Overview:
The ZDR-27 ionization vacuum gauge can be connected to one hot cathode ionization gauge, control up to four relay contact outputs, and also has functions such as analog output and serial communication. Suitable for high vacuum measurement and control.
Technical parameters:
Power supply: 220VAC/50Hz
Rated power: 10W
Operating environment: -10~40 ° C, humidity less than 85%
Control output form: Relay contact output, normally open
Relay load: 3A/220V AC, non inductive load
Instrument weight: approximately 2Kg.
Applicable regulation: ZJ-27 hot cathode ionization gauge
Measurement range and accuracy
1.0 ~ 5.0x10-5Pa, 30% of the displayed value
5.0x10-5~ 1.0x10-5, measurable
Acceleration pole voltage: 225VDC
Cathode voltage: 25VDC
Emission current: 0.1mA/1mA
Effective control range: 1.0~5.0x10-5Pa
Dimensional drawing: