1、 Main features: Measurement principle: Imported thin-film capacitive ceramic humidity sensor Measurement range: -80.0~+20 ℃ Accuracy: ± 1.5 ℃ (above
Product details


1、 Main features:

Measurement principle: Imported thin-film capacitive ceramic humidity sensor

Measurement range: -80.0~+20 ℃

Accuracy: ± 1.5 ℃ (above -60 degrees); ± 2 ℃ (below -60 degrees);

Repeatability: ≤± 1%;

320 * 240 true color TFT screen, intuitive display, Chinese and English menu interface,

Easy and convenient operation;

Selecting imported thin-film capacitive ceramic humidity sensors with long lifespan

High precision, fast response and other characteristics, can be selected according to the background gas measured on site

Different sensors;



2、 Main technical parameters

Stability: ≤± 1%/7d

Sample gas flow rate: (2 ± 0.5) L/min

Response time: T90 ≤ 3 minutes

Sample gas pressure: 0.05 MPa ≤ inlet pressure ≤ 0.25 MPa

Working power supply: 12VDC;

Dimensions: 300mm (width) x 162mm (height) x 346mm (depth)

Charging power supply: (220 ± 22) VAC, (50 ± 5) Hz,

The charger comes with charging protection function

Service life:>6 years (under normal operating conditions)

Air interface: Φ 3 stainless steel pipe (customizable size);