Z1030 Zirconia Oxygen AnalyzerDetect oxygen levels ranging from percentage to small PPM in clean gases. This analyzer is highly suitable for accurate oxygen content detection and is an important part of the original equipment manufacturer (OEM) equipment it produces. Encapsulated in a disk mounted casing, using zirconia oxygen sensors for precise detection to ensure the production of high-quality industrial gases. The unique sensor and heater design results in short preheating time, fast response, and long sensor lifespan.
Z1030 Zirconia Oxygen AnalyzerThe large multi digit digital automatic shift LCD is used to display the detected concentration and user adjustable parameters in configuration mode.
Through a fully bidirectional RS232 interface,Z1030 Zirconia Oxygen AnalyzerCan communicate with most microcontrollers. In addition to digital communication functions, it also has a fully programmable isolated 4-20mA output.
Z1030 Zirconia Oxygen AnalyzerTechnical Parameter
display
Multi bitLCD –character height12.7mm
Display Range
Display Range0.01ppm-100% O2Automatic gear shift display
display resolution
100%-103% 1%
10%-99.0% 0.1%
1%-9.99% 0.01%
0.100 -0.999% 0.001%
100ppm-999ppm 1ppm
10.0ppm-99.9ppm 0.1ppm
0.00ppm-9.99ppm 0.01ppm
Z1030 Zirconia Oxygen Analyzeraccuracy
Reading of±1%perhaps±1mVsensor output-Take the bigger one
stability
Better than full range every month2%
Sample gas flow rate
100 -300 ml/min
response speed
T-90:300ml/minUnder the traffic, it is4Within seconds
Sample gas inlet pressure (without pump installed)
10mbar-8bar
Sample gas temperature
The maximum temperature at the analyzer is100°C
Sampling system material
Stainless steel, platinum, zirconia, nickel plated brass, nylon
Sample gas inlet and outlet connection
Related to the sample gas path: Nickel plated brass (static compression seal, suitable for outer diameter of)6mm/0.25”The pipe)Double clamp brass jointDouble clamp stainless steel joint
Z1030 Zirconia Oxygen AnalyzerAnalog Output–quarantine
4 -20mA –Maximum Load500ohm
The output is programmable, and the full-scale value can be within1ppm - 100%Adjustment, the zero point value can be adjusted within0ppm- 90%Adjust.
serial communication interface
RS232DCE
Z1030 Zirconia Oxygen AnalyzerUsage environment
Temperature:0 -45 °C
Relative humidity:15 -90%Gas does not aggregate
power supply
24V DC +/- 20%, maximum current1ampere
Can be specified at the time of ordering85 - 240 Vac、50/60HzAny voltage supply between them, maximum power consumption24VA.