Advanced laboratoryBODanalyzer
YSI 5000 series laboratory dissolved oxygen analyzer combined with advanced self stirringBODAnalysis probe andYSIuniqueBODAnalysis software,
Implement laboratoryBODAnalyze automation, accuracy, and standardization.
YSI 5120typeBODanalysis software
● Installed on a computer, directly drive the dissolved oxygen meter to make the entireBODAutomation of testing process;
● The system automatically stores the readings of the first and last days, calculates and prints reports;
● Automatically read and store the dissolved oxygen reading when it reaches stability;
● Warning for supersaturated samples and temperature exceeding the range;
● Other outstanding features include: automatic blank, sample and loading compensation, non five-day testingBODResult calculation, read other dilutions on the first day
Numerical derivation, etc;
Key Features:
● traditionalBODDilution testing method,5000Series instrument coordinationYSIDissolved oxygen/BODThe probe and instrument as a whole comply with the national standard
HJ/T99-2003 standard, dissolved oxygen measurement complies with the national standard electrochemical probe methodGB11913-89Standard,BODmeasurementcomply with national standardsGB7488-87Standard;
● Advanced laboratory dissolved oxygen analyzer+self stirringBODProbe method(BODAnalysis software);
● Automatic mixingBODElectrode, can be operated with one hand;
● OptionalBODAnalysis software, implementationBODTest automation to achieve automatic data storageBODAutomatically calculate and print
report Warning for supersaturated samples and temperature exceeding the range;
● Can achieve rapid and large-scale testing;
● Strong stirring force ensures accurate test data;
● Constant stirring rate and good reproducibility;
● Built in mixer, no need for additional mixing tools, time-saving and convenient;
● use5906Type cover film: Pre installed and pre tensioned permeable film, eliminating human error caused by improper installation;
● No zero drift, no need to adjust zero;
● Membrane integrity monitoring, with automatic warning from the instrument when the membrane is damaged;
● Storable100Reading with date and time tags for the group;
Technical Specifications:
parameter |
measurement range |
resolution |
accuracy |
%Air saturation |
0to600% |
0.1% |
Accuracy ±0.1%+1lsd |
milligram/rise |
0to60milligram/rise |
0.01milligram/rise |
Accuracy ±0.1%+1lsd |
temperature |
-5to50℃ |
0.1℃ |
accurateDegree ±0.1℃ |