YHCA-100 typeStandard COD digesterFollowing the basic principles of international standards (ISO) and national standards (GB), the digestion operation with reflux heating and micro boiling for 2 hours is ensured. The preparation and addition of reagent solutions are consistent with the GB method, ensuring reliable and accurate analysis results.
The instrument adopts microcomputer technology for timed control of the heating electric furnace,Standard COD digesterMainly composed of the body, return pipeFans, electric furnace plates, etc4Most of the components are controlled by microcomputer technology for timed heating of the electric furnace plate and fan, which can10Heating a straight bottle simultaneously.To achieve the goals of energy conservation, reducing power load, and improving efficiency.
YHCA-100Type standardCODDigesterUsing glass burr reflux pipes instead of spherical reflux pipes, and replacing tap water cooling with air cooling technology. The cooling part is mainly completed by burr condensation tubes and fans. The upper part of the condensation tube is spherical, and the catalyst is added from here to prevent the instantaneous volatilization of light components in the sample. The lower part is“burr'Shape, protruding from the wall of a condenser tube on a plane2A person facing each other“Cold Bubble'Compared to a simple spherical condenser tube, it increases the cooling area and can block the passage of volatile substances and vapors. In addition, the dual effect of cooling water and internal fan in the upper spherical reflux tube ensures the reflux cooling of the sample.
The chemical solution preparation, operation, and COD calculation of the instrument shall comply with HJ 828-2017. For COD water samples below 50mg/L, the accuracy can be improved by diluting titrants and oxidants. For COD water samples above 1000mg/L, the determination can be completed by diluting the water sample proportionally.
YHCA-100Type standardCODDigesterTechnical specifications:
Measurement range: 0-1000mg/L, 0-10000mg/L (diluted water sample)
Measurement time: no more than 2 hours
Measurement error: Potassium hydrogen phthalate standard solution (500mg/L), relative standard deviation not exceeding 5.0%; Industrial organic wastewater (500mg/L), with a relative standard deviation not exceeding 8.0%
Environmental temperature: Room temperature~40 ℃
Heating power: 1500W (AC 220V ± 22V, 50HZ)
Instructions for using the instrument panel buttons:
1. Hair dryer:
Control the opening and closing of the fan.
2. Set time (timed):
Dissolve time control setting key. The instrument is set to power on for 2 hours and 30 minutes. Each press of the setting key automatically deducts 10 minutes,When using the standard reflux method to determine COD, it should be set to 2 hours.
3. Heating (start):
The digestion start control button is pressed. After pressing the 'Start' button, the instrument will perform heating digestion and automatically count down from the set value (such as 2 hours). The heating indicator light will turn on. When the set time is reached, the instrument will automatically stop heating and the heating indicator light will turn off.
Instrument usage instructions:
Before powering on the instrument, as much distilled water as possible should be added from the water injection port of the reflux pipe to ensure cooling effect.
1. Turn on the power switch on the right side panel of the instrument
At this time, the instrument displays "000", "111", "222"... "999", and finally shows the maximum set value of 2 hours and 30 minutes for digestion time. The fan is turned on at the same time, and the blowing indicator light is on.
2. Change the digestion time setting value
Press the 'Set Time' button to decrease the digestion time until the final set value is 2 hours. Of course, the digestion time value can also be set as needed.
3. Start reflux digestion
Press the 'Heat (Start)' button, and the instrument will start heating, digestion, and reflux. The heating indicator light will turn on.
4. End of digestion
After digestion, the instrument displays "000" and flashes, the heating indicator light goes out, and digestion is complete.