Adopting technologies such as photoelectric automatic detection, dot matrix graphic LCD display, touch screen buttons, etc., it has functions such as automatic recording of melting curves, automatic display of initial and final melting, etc.
The temperature control system uses precision platinum resistors as temperature detection elements, and applies digital PID regulation and PWM temperature control technology to improve the accuracy and reliability of melting point measurement. Instruments and tools
It has the function of automatically saving working parameters and storing measurement results, and can also establish communication with a PC through a USB interface or RS232 interface. The instrument adopts the hair specified in the pharmacopoeia
Thin tubes are used as sample tubes.
Main technical parameters:
Melting point measurement range: room temperature~400 ℃
Setting time for "starting temperature": 50 ℃~400 not more than 5 minutes, 400 ℃~50 ℃ not more than 7 minutes
◆ Minimum temperature display value: 0.1℃
◆ Linear heating rate: Between 0.1 ℃/min and 20.0 ℃/min, any continuous option is available
Indication error: ± 0.4 ℃ for temperatures below 200 ℃, ± 0.7 ℃ for temperatures above 200 ℃
◆ Repeatability: 0.3℃
Standard capillary dimensions: outer diameter of 1.4mm, inner diameter of 1.0mm, length of 80mm
◆ Sample filling height: 3mm
◆ Communication interface: USB or RS232 serial interface selected by touch screen buttons
◆ Data storage: 1000 sets
◆ Power supply: 220V ± 22V, 50Hz, 100W
◆ Instrument size: 400mm×280mm×230mm
◆ Instrument weight: 9.6kg