1、 Purpose and Scope of Application
The WNE-1A Engler viscometer is manufactured in accordance with the requirements of the Chinese standard GB/T266 "Determination of Engler Viscosity of Petroleum Products". Suitable for measuring the ratio of the time (in seconds) it takes for a liquid to flow out of an Engler viscometer at a certain temperature and volume to the time (in seconds) it takes for distilled water to flow out at 20 ℃, which is the Engler viscosity of the liquid, measured in conditional degrees.
2、 Main technical specifications and parameters
1. Standard water value: 51 ± 1 second;
2. Measurement temperature: 0-100 ℃;
3. Temperature control accuracy: ± 0.2 ℃;
4. En viscosity thermometer: in accordance with GB514 standard;
5. Measuring bottle specification: 200 ± 0.2ml;
6. Inner pot: made of stainless steel;
7. Electric heater power: 550W/~220V;
8. Electronic stopwatch accuracy: 1/100 second.
Shanghai Yutong Instrument Factory http://www.ytmy17.com http://www.shytyq.com/