Beiying Electronic Technology (Shanghai) Co., Ltd
Home>Products>VDL series vacuum oven
VDL series vacuum oven
The VDL series/vacuum oven with special safety performance safely dries in flammable solvents. A set of extended safety features makes drying material
Product details

VDLSeries/Vacuum Oven with Special Safety Performance

Safely dry in the presence of flammable solvents. A set of extended safety features makes drying materials containing flammable solvents very safe. For the VDL series, the TUV quality mark is a standard requirement, and the internal chamber of the equipment meets the ATEX standard; Exll-13G CIIBT3-T1 GcX. As an option, the drying oven can be upgraded to operate in Class II explosion hazardous areas in accordance with EC Directive 94/9/EG (ATEX Directive), or inert gas can be used to replace and purify areas containing circuits.

• Performance characteristics and equipment

-Electronic controlled APT.line®Internal preheating technology and two extended shelves

-Temperature range: from 15 ° C to 200 ° C above ambient temperature

-MS controller with cumulative timer (0 to 99 hours)

-Timer function of the controller: delayed opening, delayed closing, and temperature dependent delayed closing

-The slope function can be adjusted through a program editor

-Adjustable heat output (0 to 100%)

-Pressure monitor, heating release can only be activated when the pressure is less than 125 millibars

-Using compressed air to seal the instrument room against overpressure

-Fireproof lining

-Precise quantitative ventilation valve

-Adopting cross flow technology and precise inert gas quantitative supply valve

-Anti shatter safety glass panel with springs installed

-Simulated pressure gauge (displaying the pressure difference between the internal cavity and the surrounding environment)

-Inner cavity of electrolytic polishing

-All suction and ventilation pipes, as well as pressure vessels, detachable shelves, and vacuum valves welded internally, are made of stainless steel material 1.4571 (V4A)/AISI316 Ti

-The material of the door gasket is tempered silicone resin

-Independent adjustable temperature safety device, level 2 (DIN12880), equipped with a visual temperature alarm

-Rear with 16mm diameter measuring interface

-Used for communication software APT COM®Printer and communication interface RS422 for data control system


VDL23

VDL53

VDL115

Dimensions




Width (millimeters)

515

635

740

Height (including legs and casters) (millimeters)

655

775

900

Depth (millimeters)

500

545

670

Observation window width (mm)

250

350

456

Observation window height (mm)

207

307

413

Add door handles and connectors (in millimeters)

100

100

100

Distance between the rear and the wall (in millimeters)

100

100

100

Distance between side and wall (mm)

135

135

135

Internal dimensions




Width (millimeters)

285

400

506

Height (millimeters)

285

400

506

Depth (millimeters)

295

340

460

Internal volume (liters)

23

53

115

Aluminum extension shelves (standard quantity/maximum quantity)

2/4

2/5

2/6

Load per shelf (kg)

20

20

20

Allowable total load (kg)

35

45

65

Weight (empty container) (kg)

63

95

153

temperature data




Temperature range: 15 ° C to (0 ° C) above ambient temperature

200

200

200

When the temperature deviation is 100 ° C(±°C)

1.5

2

3.5

Temperature fluctuation range(±k)

0.1

0.1

0.1

Heating time to 100 ° C (minutes)

65

80

95

Vacuum interface with small flange (diameter in millimeters)

16

16

16

Test hole with small flange (diameter in millimeters)

16

16

16

Inert gas interface with flow restrictor (RP '')

3/8

3/8

3/8

Compressed air interface for sealing structures (mm)

8

8

8

Allowable maximum vacuum degree (millibars)

0.01

0.01

0.01

Vacuum leakage rate (maximum millibars/hour)

0.01

0.01

0.01

Electric data




Protection level of enclosure in accordance with EN60529 standard

IP54

IP54

IP54

Nominal voltage(±10%) 50/60 Hz (V)

230 1N~

230 1N~

230

Nominal power (watts)

800

1200

1900

Equipment safety device (A)

10

10

10

All technical data is only applicable to 22±3 ° C ambient temperature±Standard specification unloaded equipment under 10% power supply voltage fluctuation. Determine temperature data according to BINDER factory standards and DIN12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For the device series, all data are typical averages. We reserve the right to modify the technology.