VDLSeries/Vacuum Oven with Special Safety Performance
Safely dry in the presence of flammable solvents. A set of extended safety features makes drying materials containing flammable solvents very safe. For the VDL series, the TUV quality mark is a standard requirement, and the internal chamber of the equipment meets the ATEX standard; Exll-13G CIIBT3-T1 GcX. As an option, the drying oven can be upgraded to operate in Class II explosion hazardous areas in accordance with EC Directive 94/9/EG (ATEX Directive), or inert gas can be used to replace and purify areas containing circuits.
• Performance characteristics and equipment
-Electronic controlled APT.line®Internal preheating technology and two extended shelves
-Temperature range: from 15 ° C to 200 ° C above ambient temperature
-MS controller with cumulative timer (0 to 99 hours)
-Timer function of the controller: delayed opening, delayed closing, and temperature dependent delayed closing
-The slope function can be adjusted through a program editor
-Adjustable heat output (0 to 100%)
-Pressure monitor, heating release can only be activated when the pressure is less than 125 millibars
-Using compressed air to seal the instrument room against overpressure
-Fireproof lining
-Precise quantitative ventilation valve
-Adopting cross flow technology and precise inert gas quantitative supply valve
-Anti shatter safety glass panel with springs installed
-Simulated pressure gauge (displaying the pressure difference between the internal cavity and the surrounding environment)
-Inner cavity of electrolytic polishing
-All suction and ventilation pipes, as well as pressure vessels, detachable shelves, and vacuum valves welded internally, are made of stainless steel material 1.4571 (V4A)/AISI316 Ti
-The material of the door gasket is tempered silicone resin
-Independent adjustable temperature safety device, level 2 (DIN12880), equipped with a visual temperature alarm
-Rear with 16mm diameter measuring interface
-Used for communication software APT COM®Printer and communication interface RS422 for data control system
VDL23 |
VDL53 |
VDL115 |
|
Dimensions |
|||
Width (millimeters) |
515 |
635 |
740 |
Height (including legs and casters) (millimeters) |
655 |
775 |
900 |
Depth (millimeters) |
500 |
545 |
670 |
Observation window width (mm) |
250 |
350 |
456 |
Observation window height (mm) |
207 |
307 |
413 |
Add door handles and connectors (in millimeters) |
100 |
100 |
100 |
Distance between the rear and the wall (in millimeters) |
100 |
100 |
100 |
Distance between side and wall (mm) |
135 |
135 |
135 |
Internal dimensions |
|||
Width (millimeters) |
285 |
400 |
506 |
Height (millimeters) |
285 |
400 |
506 |
Depth (millimeters) |
295 |
340 |
460 |
Internal volume (liters) |
23 |
53 |
115 |
Aluminum extension shelves (standard quantity/maximum quantity) |
2/4 |
2/5 |
2/6 |
Load per shelf (kg) |
20 |
20 |
20 |
Allowable total load (kg) |
35 |
45 |
65 |
Weight (empty container) (kg) |
63 |
95 |
153 |
temperature data |
|||
Temperature range: 15 ° C to (0 ° C) above ambient temperature |
200 |
200 |
200 |
When the temperature deviation is 100 ° C(±°C) |
1.5 |
2 |
3.5 |
Temperature fluctuation range(±k) |
0.1 |
0.1 |
0.1 |
Heating time to 100 ° C (minutes) |
65 |
80 |
95 |
Vacuum interface with small flange (diameter in millimeters) |
16 |
16 |
16 |
Test hole with small flange (diameter in millimeters) |
16 |
16 |
16 |
Inert gas interface with flow restrictor (RP '') |
3/8 |
3/8 |
3/8 |
Compressed air interface for sealing structures (mm) |
8 |
8 |
8 |
Allowable maximum vacuum degree (millibars) |
0.01 |
0.01 |
0.01 |
Vacuum leakage rate (maximum millibars/hour) |
0.01 |
0.01 |
0.01 |
Electric data |
|||
Protection level of enclosure in accordance with EN60529 standard |
IP54 |
IP54 |
IP54 |
Nominal voltage(±10%) 50/60 Hz (V) |
230 1N~ |
230 1N~ |
230 |
Nominal power (watts) |
800 |
1200 |
1900 |
Equipment safety device (A) |
10 |
10 |
10 |
All technical data is only applicable to 22±3 ° C ambient temperature±Standard specification unloaded equipment under 10% power supply voltage fluctuation. Determine temperature data according to BINDER factory standards and DIN12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For the device series, all data are typical averages. We reserve the right to modify the technology.