Applicable standards: GB/T11133 GB/T11146 GB/T7600 GB/T6023 GB/T6283 GB/T606
The SCKF102 micro moisture analyzer adopts the classical theory of Karl Fischer microcoulometry; According to the law of electrolysis, the number of water molecules reacted is directly proportional to the number of charges. The instrument detects the number of charges involved in the reaction (Coulomb) and automatically converts it into the corresponding number of water molecules. Therefore, this method has extremely high testing accuracy, extremely low testing cost, and irreplaceable advantages over other testing methods; Can reliably determine trace amounts of moisture in liquid, gas, and solid samples. Featuring high sensitivity, high precision, high reproducibility, and low-power energy-saving design, it is widely applicable in fields such as petroleum, chemical, power, pharmaceuticals, commercial inspection, scientific research, and environmental protection.
The types of detectable substances include:
1. Gasoline, hydraulic oil, insulation oil, transformer oil, turbine oil, fire-resistant oil.
2. Pentane, hexane, dimethylbutane, octane, dodecane, eicosane, octadecane, cyclododecane, decylcyclohexane, methylbutadiene, benzene, toluene, xylene, ethyltoluene, dimethylstyrene, tetracene, petroleum ether, cyclohexylamine, methylcyclohexylamine, cyclohexane, ethylcyclohexylamine, dicyclopentadiene, dimethylnaphthalene, trimethylstyrene, biphenyl, dihydrophenanthrene, fluorene, methylethylphenanthrene, isopropylbenzene, etc.
3. Phenolic phenols, cresols, fluorophenols, chlorophenols, dichlorophenols, nitrophenols, etc
4. Ether type diethyl ether, diethylene glycol monomethyl ether, diethylene glycol diethyl ether, polyethylene glycol diethyl ether, benzyl ether, fluorophenyl ether, iodobenzene ether, didecyl ether, and diheptyl ether
5. All alcohols, all halogenated hydrocarbons, all lipids
instrumentcharacteristic
128 × 64 dot matrix LCD display screen, no labeled button operation;
Use blank current compensation and balance point drift compensation to correct measurement results;
Switching constant current electrolysis technology reduces the overall power consumption of the machine;
◆ Historical records with time stamps, up to 500 can be stored;
◆ Equipped with electrode open circuit and short circuit self-test alarm function;
◆ Equipped with screen saver function to extend the lifespan of LCD;
◆ Can achieve full Chinese/full English interface display
Technical Parameter
Measurement range: 0ug~200mg
Accuracy: The tested water volume is between 3ug and 1000ug, with an error of less than ± 3ug
The test water volume is greater than 1000ug, with an error of less than ± 0.3%
Resolution: 0.1ug
Electrolytic current: 0-400mA
◆ Standby power consumption: 4W Maximum power consumption: 20W
Power supply voltage: AC220V ± 20% 50HZ ± 10%
◆ Applicable ambient temperature: 5 ℃~40 ℃
◆ Applicable environmental humidity: ≤ 85% RH
◆ Dimensions: 270 × 210 × 125 (mm)
Weight: Approximately 3kg
Micro moisture analyzerSCKF102typePatent certificate