Total phosphorus measuring instrument
The German WTW total phosphorus measuring instrument adopts the molybdenum blue colorimetric method, which has two-point calibration, automatic digest
Product details
GermanyWTWTotal phosphorus measuring instrumentTresCon TP
For applications that require accurate testing and real-time adjustment of water treatment processes,TresConIt was their wise choice. Compared to on-site direct input sensors,TresConIt has the characteristics of more accurate and stable testing, and stronger anti-interference.TresConAdopting modular design, * allowing for more3The arbitrary combination of modules and multi parameter monitoring system have unparalleled cost advantages.

Testing Principles
The total phosphorus module is divided into two parts, the * * part is a high-temperature chemical digestion module used to heat and digest samples, and the * * part is an analysis module used to test phosphate concentration.
During the digestion process, all valence states of phosphorus compounds are converted into orthophosphate that can be analyzed by photometry. The reaction is based on the oxidation of other valence states of phosphorus to orthophosphate by persulfate under acidic conditionsPO43-High temperature and high pressure conditions can greatly shorten the digestion time.
The molybdenum blue colorimetric method is used to test the concentration of orthophosphate: molybdate is added to the sample, and a chemical reaction is carried out to make the sample blue. The color change is sensed by a photometer and then converted into concentration. The entire process is carried out in an orderly manner under the control of the controller, thus ensuring high operational safety.
Total phosphorus measuring instrumentTresCon TPTechnical Parameter
Sample pretreatment
A separate sample preprocessor is required, such asPurconperhapsPF105
Sample supply
Equipped with a standard flow channel, the overflow sample volume is sufficient3An analysis module can be used, and it can also be installed simultaneously3A circulation channel is used for testing3A different testing point
Outputs
3Group isolation0/4-20 mAOutput,12Passive relay output,RS232andRS485
power supply
230 VAC ± 10%, 50 Hz / 115 VAC ± 10%, 50 – 60 Hz
Surrounding environment
Storage temperature:25 . .. 60 °CWorking temperature:0 .. 40 °C
authentication
CE, DIN-GOST
Protection level
IP54IEC 1010-1/EN 61010-1, Class 1
weight
27 kg(Excluding analysis module)Each analysis module is reconfigured10 kg,Installation bracket weight approximately25 kg
TresConAutomation function
AutoClean
Automatic cleaning function, tested the sewage plant discharge outlet without any problems
AutoCal
Automatic calibration function, adjustable calibration interval, ensuring better testing accuracy
AutoKorr
Automatic background value correction function to improve the accuracy of photometric analysis
AutoFlow
Automatic container liquid level monitoring function, automatically prompts for filling reagents or pipeline blockage
AuthTherm
Automatic constant temperature function
Test range
Total phosphorus:0.01 …3.00 mg/l, 0.3 … 100 µmol/l
0.01 … 6.00 mg/l, 0.3 …
200 µmol/l (1:1dilute)
resolution
Test range:0.01 …3.00 mg/l: 0.01 mg/l
0.30 …
100 µmol/l: 0.1 µmol/l
accuracy
±3%testValue ±0.05 mg/l P
measuring principle
Reference beam colorimetric method after digestion
test method
Molybdenum blue colorimetric method
Method deviation coefficient
1.5%
reaction time
4minute
Test Interval
10,15,20,25,30,60Minutes optional; DINDissolving method(120 °C, thirtyMinutes)
correct
automatic2Point Correction
Sample consumption and requirements
10 ml/order,Suspended particles<50 mg/l
Reagent consumption
Reaction reagents(A,B,C,D)Available60Every day (every60Measured every minute),
Standard solution and cleaning solution are available60order
reagentABCD10/15/20/30/60Oh my god, every10/15/20/30/60Minute measurement interval
Standard liquid1.5 l70Oh my god,24Hourly calibration interval
cleaning solution1.5 l60Oh my god,24Hourly cleaning interval
service interval
every3Month