Equipment Overview:
TY-5006Hot deformation Vicat softening point testerDetermine the thermal deformation and Vicat softening point temperature of thermoplastic materials and thermoplastic pipe fittings.
Compliant with standards:
1. GB/T 8802-2001Determination of Vicat softening temperature of thermoplastic pipes and fittings
2. GB/T 1633-2000Determination of Vicat Softening Temperature (VST) of Thermoplastic Materials
3. GB/T 1634.1-2004Determination of plastic deformation temperature under load - Part 1: General test method
4. GB/T 1634.2-2004Determination of Load Deformation Temperature of Plastics Part 2: Plastics, Hard Rubber, and Long Fiber Reinforced Composite Materials
5. GB/T 1634.3-2004Determination of plastic deformation temperature under load - Part 3: High strength thermosetting laminates
6. ASTM D648-01《Standard Test Method for Deflection Temperature of Plastics Under Flexural Load in the Edgewise Position》
Performance characteristics:
1. Using imported high-precision digital dial gauges as sample deformation measuring elements, with high accuracy and good stability, fully ensuring the accuracy of test data;
2. The connecting rod of the sample holder and the load rod are made of low expansion alloy with the same diameter and material, which can minimize the deformation of the sample holder and improve the accuracy of the test results;
3. Equipped with automatic lifting function for the sample rack, driven by a linear motor with strong load-bearing capacity and smooth operation;
4. Can simultaneously perform two types of tests: thermal deformation and Vicat softening point temperature;
5. There are water cooling, natural cooling and other methods to choose from according to user needs;
6. Equipped with multiple safety protection devices to ensure the safety of testing equipment and operators;
7. Specialized experimental software with powerful functions, user-friendly interface, and easy operation.
Internal structure:
Technical parameters:
Model |
TY-5006A |
TY-5006B |
TY-5006C |
Temperature control range |
Room temperature -300 ℃ |
||
Temperature Accuracy |
±0.5℃ |
||
Uniform heating rate |
ASpeed: 5 ± 0.5 ℃/6 min B Speed: 12 ± 1.0 ℃/6 min |
||
Using media |
Silicone oil or transformer oil (to be provided by the user) |
||
Deformation measurement range |
>10mm |
||
Deformation resolution |
0.001mm |
||
Number of sample racks |
3 |
4 |
6 |
Load |
(1) Vika load: GA=10N+± 0.2N GB=50 ± 1N (2) Hot deformation bending normal stress: Method A, using fiber stress of 1.80Mpa Method B, using fiber stress of 0.45Mpa Method C, using fiber stress of 8.00Mpa |
||
Maximum heating power |
≤4500W |
||
Select accessories |
Software and computer |
||
power supply |
AC three-phase five wire 380V/50Hz |
||
Cooling method |
Natural cooling and water cooling |
||
Special Features |
Equipped with upper limit temperature protection function setting and can be equipped with an oil suction smoke system |
||
size |
1055×750×760mm |
1255×750×760mm |
1355×750×760mm |
weight |
180kg |
220kg |
260kg |