THU1010 eraUltrasonic hardness tester
Product Overview:
There are various methods currently available for testing hardnessCommon testing methods include Brinell, Rockwell, Vickers, and Richter tests. Among them, the Brinell and Rockwell tests have high force and large indentation, causing significant damage to the surface of the specimen. Vickers uses optical methods for measurement, which requires very professional technicians to operate and cannot directly test large workpiecesBeijing EraTHU1010 eraUltrasonic hardness testerThe comparative measurement of samples using ultrasonic contact impedance method has the advantages of high accuracy, fast speed, easy portability, simple operation, and non-destructive measurement.
THU1010 eraUltrasonic hardness testerFunctional Features
● High measurement accuracy——accessible±3%HV,±3%HB,±1.5HR.(This measurement accuracy was measured using a test bracket)
● Small test indentation——It requires a high-power microscope to observe.
● Fast measurement speed——Can be in2Output test results within seconds.
● Large screen display——Directly display the current measurement value, cumulative measurement value, * maximum value, * minimum value, average value, and unit conversion value.
● Massive storage——Can be stored in the instrument's internal memory2000Group measurement data.
● Simple calibration——Can be stored in the instrument's internal memory20Group correction data, oneGroup default data,Users can set19groupCall anytime.
THU1010 eraUltrasonic hardness testerApplication scope
● Hardness measurement of flange edge and gear root stamping parts, tooling, thin plates, surface hardened teeth and gear grooves, and tapered parts;
● Hardness measurement of shafts, thin-walled pipelines, and containers;
● Hardness measurement of wheels and turbine rotors;
● Hardness measurement of drill bit edge,Hardness measurement of machine tool cutting edge and tool holder;
● Hardness measurement of welding parts;
● Measure deep holes with a certain aperture depth, dents and protrusions with large curvature, irregular planes;
● Can cover the hardness measurement of the vast majority of black metals, non-ferrous metals, and their alloys in industrial production
●Hardness measurement can be performed on surface electroplating layers.
THU1010 eraUltrasonic hardness testertechnical parameter
● Test scope:
HV:50~1599; HRC: 20.3~76; HB: 76~618;
HRB :41~100; HRA:61-85.6; Mpa: 255~2180N/mm2
● Test direction: Supports 360 °As long as the pressure head is aligned with the measured surface90°±5°The angle can be used for measurement operations.
● Hardness value standard: Vickers (HV)),Roche C(HRC),Brinell (HB))Wait.
● Battery specification: 4.2V,4800mAhLithium batteries.
● Battery life: 10Hour.
● It can be connected to an external portable printer to print measurement data or to a computer super terminal to export measurement data
● Can store 2000Group measurement data and20A calibration data set.
● The average value can be calculated by accumulating multiple measurements on the tested workpiece with uneven distribution of hardness values.
● The optional testing bracket has fast batch testing speed, easy operation, and high testing accuracy for small finished workpieces.
● No damage to the surface of the tested workpiece, good stability, and high testing accuracy.
● Ultrasonic hardness tester host specification: 162 × 81 × 31mm
● Probe specification: Φ 22 × 153mm
● Ultrasonic host (including battery) and probe (HP-2K))Total weight:755g.
Standard configuration:
HostHP-2KManual probe, hardness block, charger, data cable, instrument box, file
Optional attachments:
1/2/5/10KgManual probe0.3/0.5/0.8KgElectric probe, test stand, printer, ultrasonic Rockwell hardness block, ultrasonic Vickers hardness block
The selection criteria for probes are shown in the table below(The actual indentation depth is approximatelylowerDescribe the data10%):
|
Ultrasonic hardness testerDifferent probesmeasureThe small thickness corresponding to a specific hardness value(Company:μm)(10Multiple indentation depth) |
||||
|
|
HP-1K |
HP-2K |
HP-5K |
HP-10K |
|
800HV |
69 |
100 |
150 |
220 |
|
600HV |
79 |
110 |
180 |
250 |
|
300HV |
112 |
160 |
250 |
350 |
|
Probe type |
1kg manual probe HP-1K (optional) |
2Kg manual probe HP-2K (standard) |
5Kg manual probe HP-5K (optional) |
10Kg manual probe HP-10K (optional) |
|
Real testing power |
10N |
20N |
50N |
98N |
|
diameter |
22mm |
22mm |
22mm |
22mm |
|
length |
154mm |
154mm |
154mm |
154mm |
|
Resonant rod diameter |
2.4mm |
2.4mm |
3mm |
3mm |
|
Measurement surface * large roughness |
Ra<3.2um |
Ra<5um |
Ra<10um |
Ra<15um |
|
Workpiece * small weight |
0.3kg |
0.3kg |
0.3kg |
0.3kg |
|
Workpiece * small thickness |
2mm |
2mm |
2mm |
2mm |
|
Application scenarios of measuring heads |
Ion nitriding die, mold shell, fixture, thin-walled part, bearing, tooth side and pipeline inner wall |
Ion nitriding die, mold shell, fixture, thin-walled part, bearing, tooth side and pipeline inner wall |
Measure grooves, tooth sides, and tooth roots |
Small forgings, casting materials, weld inspection, heat affected zone, with low requirements for roughness |