Sujing Antai Chengdu AgentChengdu Chuanhongke Biotechnology Co., Ltd. sincerely provides services for colleges and universities, medical and health care, pharmaceutical companies, and research institutes and laboratoriesSujing Antai BSC-1604IIB2 Level 2 Full Row Biosafety Cabinet Modelas well asSujing Antai BSC-1604IIB2 Level 2 Full Emission Biosafety PriceProfessional technical engineers promise that all Sujing Antai BSC-1604IIB2 Level 2 Full Row Biosafety Cabinets have three guarantees of quality, one-year warranty, lifelong maintenance, easy installation and commissioning, and technical training!
Sujing AntaiBSC-1604IIB2Biosafety cabinetProduct features:
1. Vertical laminar negative pressure machine. The air is filtered and discharged outdoors or connected to the exhaust system.
2. High quality HEPA or ULPA filters with "zero leakage" patented technology are installed in both the air supply and exhaust systems to ensure cleanliness at ISO 5 (level 100) or ISO 4 (level 10). (Patent No. ZL.2)
3. A double-layer box surrounded by negative pressure ensures no pollution or leakage. All work areas are made of SS304 stainless steel, and the rounded corner inner liner is molded in one go to increase self-cleaning function.
4. Configure SS304 all stainless steel collection tank with a capacity greater than 4 liters.
5. The 10 ° inclined design of the isolation operation surface is more in line with the principles of ergonomics, making the operator more comfortable.
6. The sliding front window adopts a suspension lifting system imported from Japan, which can be adjusted and positioned freely with safety glass thicker than 5mm, reliable performance, and maintenance free. Close the seal for easy sterilization treatment. (Patent No. ZL.5)
7. The front suction inlet adopts patented technology for unobstructed return air. (Patent No. ZL.8)
8. Detachable arc-shaped hand rest board reduces work fatigue and is easy to transport.
9. A fully Chinese human-machine dialogue interface, operated with a touch of a button. The LCD screen displays real-time parameters such as descending wind speed, suction inlet wind speed, filter service life and blockage alarm, fan operation status and fault alarm, and real-time monitoring and display of unit operation time.
10. The front window opening height limit sound and light alarm system is linked with the lighting control.
11. Safety interlock system for lighting and sterilization systems.
12. The workspace is equipped with splash proof safety power sockets and various gas pipe connection valves for easy use by operators.
13. The imported fan intelligent air volume automatic compensation system ensures that the air volume change of the fan is less than 10% when the filter resistance increases by 50%, improving safety.
14. The air supply and exhaust pipelines adopt BELIMO actuator PDI control valves, effectively reducing the impact of external airflow on the safety cabinet.
15. Strict cabinet leak prevention testing ensures that the cabinet has no leaks under 500Pa conditions.
16. Strict HEPA/ULPA filter leak detection ensures that the leakage rate of scannable filters is ≤ 0.01%, and the leakage rate of non scannable filters is ≤ 0.005%.
Sujing AntaiBSC-1604IIB2Technical parameters of biosafety cabinet:
Air cleanliness: ULPA: ISO4 level
Filter level/filtration efficiency: ULPA ≥ 99.9995%, @ 0.12um
Descending wind speed (m/s): 0.35
Inflow wind speed (m/s): 0.55
Noise (dB (A)): ≤ 65
Power supply: AC220V 50Hz
Airflow balance biological protection: 1. Cross infection protection: A The total bacterial count of the impact sampler is ≤ 10CFU per time
The total bacterial count of the B slit sampler is ≤ 5CFU per time
2. Personnel protection: total bacterial count ≤ 5CFU./time
3. Protection of tested products: total bacterial count ≤ 2CFU./time
Rated power (including spare socket) (Kw): 1.75
Weight (Kg): 370
Work area size (mm) (W * D * H): 1604 * 630 * 630
Dimensions (mm) (W * D * H): 1800 * 795 * 2050
Lighting specifications (W) and quantity: 36 * ①
UV lamp specifications (W) and quantity: 40 * ①
Illuminance (Lx): ≥ 900
Exhaust direction: Top out