Demada Precision Instrument Co., Ltd
Home>Products>Standard type (0.1um resolution) LGF series probe
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    13962107506
  • Address
    Room B-401, Chengjian Building, 185 Jumao Road, Xiangcheng District, Suzhou
Contact Now
Standard type (0.1um resolution) LGF series probe
Standard type (0.1um resolution) LGF series probe
Product details

542-181/182 Sanfeng Standard LGF Linear Micrometer

542 series grating micrometer - standard design

characteristic:

The resolution of the LGF series micrometer is 0.1 μ m.
2. Factory workshop. Excellent dust/water protection level reaches (IP66) protection level.


Technical parameters:
Accuracy: Refer to performance parameters (excluding quantization bias)
resolving power: 0.1μm
Length benchmark: Transmission type photoelectric linear encoder
Maximum response speed: 400mm/s
Measuring needle: ø 3mm hard alloy
axle sleeve: ø 8mm or ø 15mm
Bearing type: Linear ball bearing
Force measurement: refer to performance parameters
Output signal: 90 degree phase difference, differential square wave (RS-422A equivalent)
Signal spacing: 0.4μm
Dust/water protection level: up to IP66 protection standard


Performance parameters:

Item number

542-181

542-182

measuring range

10mm

25mm

Resolution

0.1µm

Measurement accuracy (20 º C)

(0.8+L/50) µ m (L=any measured length (mm))

Quantitative bias

±1

Force measurement

The measuring rod is facing upwards

Below 1.0N

Below 4.0N

Horizontal measuring rod

Below 1.1N

Below 4.3N

The measuring rod is facing downwards

Below 1.2N

Below 4.6N

Length benchmark

Photoelectric linear encoder

Response speed * 1

400mm/s

output signal

90 º phase difference, differential square wave (RS-422A equivalent) with minimal edge spacing, 200ns

Output signal spacing

0.4µm

weight

About 310g

About 350g

Dust/water protection level * 2

Achieve IP66 protection level (only for micrometer)

Measuring head

ø 3mm hard alloy (fixing screw: M2.5 (P=0.45) × 5) standard probe No.901312

Clamp sleeve

ø8mm

ø15

Bearing type

Linear ball bearing

Output cable length

2m (extending directly from the micrometer device)

Connector

Plug: RM12BPE-6PH ( HIROSE ) ,

Compatible sockets: RM12BRD-6S ( HIROSE)

Operating temperature (humidity) range

0-40 º C (RH 20-80%, no condensation)

Storage temperature (humidity) range

-10-60 º C (RH 20-80%, no condensation)

Standard attachments

Special wrench for measuring head: No.538610

Special wrench for measuring head: No.210187




Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!