Product Introduction
1、 Introduction
Spee Super Oil Condition Monitoring InstrumentFluidScan 1100 is easy to operate, only one drop of sample is needed, and the detection result can be obtained in 1 minute. The device includes an expandable oil database, allowing users to easily add oil samples from outside the database. This device is based on direct reading infrared spectroscopy (DIR) technology, which is used to directly quantitatively analyze the key state indicators of lubricating oil, and directly measure the aging and pollution levels of synthetic oil or mineral oil. The direct quantitative detection results include: total acid value TAN, total base number TBN, oxidation degree, nitrification degree, sulfurization degree, additive loss, mixed oil pollution, micro water, residual carbon, ethylene glycol (coolant pollution), oil matching degree, and fatty acid methyl ester (FAME) in biodiesel.
IISpee Super Oil Condition Monitoring Instrumentmain features
1. Adopting solid-state optical waveguide technology;
2. High accuracy of detection results;
3. Rich industrial oil quantitative analysis database;
4. No solvent required, no oil sample preparation required, minimal oil sample consumption (100ul);
5. Built in Chinese and English system. Provide desktop management software;
6. The testing cost is very low;
7. Lightweight, easy to carry, 6-8 hours of battery life;
8. Meet the requirements of ASTM D7889-13 standard;
9. The measurement accuracy of Total Acid Number (TAN) and Total Base Number (TBN) is comparable to the detection accuracy of ASTM D4739 and ASTM D664 (Oil Determination Testing Standards);
10. The measurement range of moisture content has been expanded to 6.5% (included in all MiniLab systems).
3、 Main parameters
(1)Product application:
1. Application scope: Mineral or synthetic lubricants and biodiesel used in gearboxes, engines, transformers, hydraulic systems, turbines;
2. Output result: TAN(mgKOH/g)、TBN(mgKOH/g)、 Oxidation degree (abs/0.1mm), nitrification degree (abs/0.1mm), vulcanization degree (abs/0.1mm), micro water (ppm), ethylene glycol (w%), soot (w%), mixed oil pollution (w%), loss of antioxidant additives (% residue), loss of anti-wear additives (w%);
3. Standards used: ASTM D7889, E2412 (revised);
4. Infrared in the wavenumber range: 900-4000cm-1;
5. Accuracy: ≤± 3% of measured values (typical);
6. Repetition accuracy: ≤± 6% of measured values (typical)‘
7. Calibration: No need.
(2) Using parameters
1. Injection volume: 0.03 μ L (approximately 1-2 drops);
2. Solvent/reagent: No need;
3. Working temperature:- 10℃-50℃;
4. Working humidity: 0-100%, no condensation;
5. Working altitude: ≤ 5000 meters.
(3)Power supply parameters
1. Battery: The power supply is equipped with a built-in rechargeable lithium-ion battery;
2. Power: AC 240V, 50Hz, 10W;
3. Typical working hours: 6-8 hours;
4. Charging time approximately: 6.5 hours;
(4) User interface parameters
1. Software/Operating System: Microsoft Windows ® CE;
2. Display: 320 × 320 reflective color display screen;
3. Data storage: up to 5000 analysis results;
4. Data transmission: USB;
5. Data input: directional keys and soft buttons.
(5) Mechanical parameters
1. Size: 240mm (height) x 140mm (width) x 70mm (length);
2. Weight: Approximately 1.4 kilograms;
3. Outer packaging size: 17.1cm (height) x 6.3cm (width) x 5.5cm (length);
4. Color: Blue
(VI)Consumables
1 PV1012: Disposable straw (60 μ L) and soft cleaning paper pack, 100 sets/pack;
2. FL310: IR verification oil 5mL;
3 PV1011: Disposable soft cleaning paper, 500 sheets/pack.