Solid particle power frequency dielectric loss dielectric constant tester
Meet the standards:
GB/T1409-2006 Recommended methods for measuring the permittivity and dielectric loss factor of electrical insulation materials at power frequency, audio frequency, and high frequency
GB/T 5654-2007 Measurement of relative permittivity, dielectric loss factor, and DC resistivity of liquid insulating materials
GB/T 21216-2007 Test Method for Measuring Conductivity and Capacitance of Insulating Liquids to Determine Dielectric Loss Factor
GB/T 1693-2007 Determination method for dielectric constant and dielectric loss tangent of vulcanized rubber
GB/T 5594.4-1985 Test Method for Properties of Ceramic Materials in Electronic Component Structures Test Method for Tangent Value of Dielectric Loss Angle
……………………………………………………………………………………………
1、 Product Overview
Solid particle power frequency dielectric loss dielectric constant tester, is an advanced instrument for measuring dielectric loss (tg δ) and capacitance capacity (Cx), which measures the dielectric loss (tg δ) and capacitance capacity (Cx) of various insulation materials, insulation sleeves, insulation liquids, power cables, capacitors, transformers, and other high-voltage equipment. It has the advantages of simple operation, Chinese display, printing, convenient use, no need for conversion, built-in high voltage, strong anti-interference ability, and short testing time.
This tester adopts variable frequency power supply technology and utilizes microcontroller and electronic technology for automatic frequency conversion, analog-to-digital conversion, and data operation, achieving strong anti-interference ability, fast testing speed, high accuracy, and easy operation.
2、 Performance characteristics
1. The instrument has high measurement accuracy and can meet the requirements of oil dielectric loss measurement. Therefore, only a standard oil cup and a dedicated testing line are needed to achieve oil dielectric loss measurement.
2. By using frequency conversion technology to eliminate on-site 50Hz power frequency interference, reliable data can be measured even in strong electromagnetic interference environments.
3. The overcurrent protection function ensures that the instrument is not damaged in the event of a short circuit or breakdown of the test sample.
4. Comes with standard capacitors and high-voltage power supply for easy on-site testing and reduces on-site wiring.
5. The instrument adopts a large screen LCD display, and the testing process is both intuitive and easy to operate through Chinese character menu prompts.
3、 Technical indicators
technical indicators
1. Test environment temperature: 10 ℃~30 ℃ (LCD screens should avoid prolonged sunlight exposure)
2. Relative humidity: 20% to 80%
3. Power supply: Voltage: 220V ± 10%
4. Dimensions: Length * Width * Height=470mm * 320mm * 360mm
5. Weight: 16kg
6. Output power: 1.5KVA
7. Display resolution: 3 bits, 4 bits (all internal are 6 bits)
8. Testing methods: forward connection method, reverse connection method, external test voltage method
9. Measurement range: Internal test voltage:
tgδ:99.9%
Cx :30 pF<Cx(10KV)<60000 pF
10KV Cx<60000 pF
5KV Cx<80000 pF
2.5KV Cx<0.3uF
External test voltage:
Determined by the output power of the external test transformer
10. Basic measurement error: dielectric loss (tg δ): 1% ± 0.09%
Capacitor capacity (Cx): 1.5% ± 1pF
11. Resolution: tg δ: 0.01%
Cx :0.1pF
12. The instrument has an integrated structure for boosting and measuring, with five adjustable output voltages ranging from 2.5KV to 10KV
13. Sample requirements: Diameter of 50MM, 100MM, 38MM