1 Small sized atomized hydrogen peroxide sterilizerProduct Introduction
The space atomization hydrogen peroxide sterilization system uses the gasification method of spore killing agent dry mist to sterilize the overall space and surface of the clean area. Through professional sterilization schemes and software, it provides a complete set of thorough, convenient, safe, environmentally friendly and reliable best methods for sterilization of sterile clean areas for pharmaceutical customers. At the same time, challenging experiments on bacterial spore biological indicators can reach 10%3~106The secondary killing effect. The biggest features of the atomized hydrogen peroxide sterilization system are: good sterilization effect, short operation time, no residual sporicidal agent, non corrosive environment, low operating cost, and complete validation support.
II working principle
The spray process is completed by diffusing dry fog to the area to be sterilized. When the average diameter of droplets is less than 10 microns, the sprayed mist can be referred to as "dry"; Small droplets will bounce off the wall and will not break or adhere, making the surface moist. And all the conditions are established to meet the requirement that fungicides can effectively move to designated areas in the form of atomization, which determines that they can move to areas that are difficult to reach normally.
Dry fog has the following properties:
Dry droplets do not settle and undergo irregular motion (Brownian motion principle);
Dry droplets will not aggregate together to produce large droplets;
Dry droplets will rebound upon surface contact without breaking, thus wetting the surface;
Therefore, these properties of dry mist gas provide good spatial and surface contact effects in areas that are difficult to reach.
![]() |
|||
![]() |
|||
Small droplets will bounce back, so they won't break. Large droplets break, so they adhere wet
III Scope of Application
Suitable for places that require regular disinfection and sterilization, such as biosafety laboratories, animal centers, and infectious disease wards.
4 Technical Features
uCompact in size and simple in structure, suitable for use in pharmaceutical company laboratories, sealing rooms, and material warehouses;
uColor LCD touch screen control display, device status display is more intuitive;
uConfigure wireless remote control to remotely control device startup and pause;
uCompliant with pharmacopoeiaGMPAccording to regulatory requirements, thermophilic Bacillus subtilis is used as the challenge bacterium to achieve cleanliness in the clean area6Logarithmic killing rate;
uA single machine can be used for20-250mDisinfection of the space;
uPermission management, password function;
uEquipped with delayed start function;
5 Technical Specifications
Power supply: AC220V/50Hz
Power: 180W
Spray speed: 15ml/min
Working hours: ≥ 3H
Sterilization agent: liquid sporicidal agent
Sterilization volume: 20-250m ³
Killing rate: The ability to kill thermophilic fat spores is 6-log
Control mode: wired or wireless control
Size: 350 * 260 * 280mm
Weight: 13kg
VI Structural Description

A——Solution chamber plugB——Atomizing nozzleC——ChassisD——Control panelE——Antenna plug F——Power switchG——Power socketH——Drain nozzle
7 Operation steps

8 Device verification
u Provide a comprehensive validation plan, including IQOQPQ, in compliance with the new GMP requirements.
u Fully compliant with the regulations of the United States Pharmacopeia (USP), using Bacillus subtilis or thermophilic Bacillus subtilis as the challenge test bacteria, achieving a killing rate of 3-6 logarithmic units from 10 ^ 3 to 10 ^ 6, which proves the system's ability to kill various microorganisms in the process of space sterilization and purification.
u After the sterilization space air conditioning system is fully ventilated for 30-120min, the residual hydrogen peroxide is less than 1PPm and the residual acetic acid is less than 10PPm, which meets the requirements of OSHA (Occupational Safety and Health Standards) in the United States.
9 Comparison of sterilization methods
project |
Atomized hydrogen peroxide Sterilization system |
Formaldehyde fumigation |
ozone |
residue |
Basically, there is zero residual surface space |
It is necessary to clean the white crystals after the sterilization process, and it is difficult to completely remove any residue |
It is necessary to remove residues, and it is equally difficult to completely remove them |
|
Chemical toxicity |
Hydrogen peroxide is non-toxic |
Carcinogenic |
Strong oxidizing properties and hazards to the human respiratory system,EPANot certified |
|
Disinfection and sterilization effect |
㏒4~㏒6Broad spectrum sterilization effect, such as spores, fungi, etc |
Cannot fully achieve ㏒4~㏒6The spore killing effect |
At high concentrations, it has the effect of killing spores, but due to its fast lysis, it is difficult to achieve the effect of killing spores in a clean room |
Overall sterilization time |
2~6hour |
at least2sky |
The majority of users operate on a daily basis, which poses a potential risk to the compatibility of the cleanroom environment |
Residual neutralization |
No need for neutralization |
Need to neutralize and clean |
———— |
certification |
The atomized hydrogen peroxide sterilization system passes throughEPAverify |
Advanced pharmaceutical and biotechnology companies in Europe and America have abandoned this sterilization method |
noneEPAverify |
Residual detection system |
By using a spatial residual detector, residual hydrogen peroxide can be detected |
———— |
———— |
Cost evaluation |
From residues, chemical toxicity, sterilization effectiveness, and compliance with national standardsGMPVarious aspects such as certification of national and international standards need to be considered in terms of cost. Less downtime, better employee protection, and easy operation all determine the high cost-effectiveness of the atomized hydrogen peroxide sterilization system. |
||