Shenzhen York Technology Co., Ltd
Home>Products>Six function verification furnace V2140
Six function verification furnace V2140
Product Name: Six Function Calibration Furnace V2140 Product Model: Six Function Calibration Furnace V2140 Product Exhibitor: Shenzhen York Technology
Product details
Six function verification furnace V2140 Detailed Introduction
Venus 2140 provides an additional wide operating range, which is easy to use and convenient to carry. As one of the ISOCAL-6 series, V2140 can be used as a dry body furnace, liquid tank, blackbody source for infrared thermometers, surface temperature sensor calibrator, and several m K (0.001 ℃) ITS-90 fixed point calibrators. There are two models of Venus 2140, Basic (Type B) and On site (Type S). The B-type comes with a sophisticated temperature controller that displays the set temperature and dry body temperature. The S-type has a built-in digital indicator that can be connected to an external temperature sensor, providing better accuracy and traceability. When using the dry body furnace function, we recommend choosing the on Site dry body calibration furnace and selecting the on-site precision temperature probe. It can be used as an external standard for upper level traceability and calibration to establish an on-site temperature calibration laboratory. The UK based company Assunt uses an external probe connected to the on-site indicator to compensate for temperature gradients and load errors. We recommend the option of embedded independent indicator (field type) and standard platinum resistance thermometer to make it a completely independent calibration laboratory. Using the fixed point method, the calibration slot provides unprecedented accuracy of ± 0.001 ℃; By using a stirred tank method, the accuracy can reach ± 0.005 ℃ (compared to calibration)



Indicator V2140 type
Temperature range: Below the ambient temperature of 55 ℃~+140 ℃
Absolute minimum temperature: -55℃
Absolute stability exceeding 30 minutes
1 Dry body verification furnace: ±0.03℃
2 Mixing liquid tank: ±0.025℃
3 Mixing ice/water tank: ±0.001℃
4 Blackbody Furnace: ±0.3℃
5 Surface temperature sensor calibrator: ±0.5℃
6 1TS-90 fixed point equipment: ±0.0005℃
Stable time: 10 minutes
Verification capacity: 35mm diameter x 160mm depth
uniformity: ±0.018℃
Controller resolution: 0.1℃
Indicator resolution: 0.1℃ ~ 0.01℃
Indicator unit: ℃,℉, K
External dimensions: 302mm(H)×176mm(W)×262mm(D)
Weight: 10.2kg

Ordering reference
Fortunately, users do not need to purchase the entire verification package at once. For example, you can purchase a dry body calibration furnace first and add accessories according to your own needs in the future. Due to the replaceability of various attachments, ISOCAL-6 has greater flexibility.
option function model
1 Dry body verification furnace V2140B
2 Mixing liquid tank Basic type with temperature controller
3 Mixing ice/sink V2140S
4 Black body furnace On site type with temperature controller and indicator
5 Surface thermometer calibration furnace V2140SN
6 ITS-90 fixed point equipment On site type, equipped with temperature controller, indicator, NAMAS certificate and standard PT100 probe

Options
1. User customized casing or replacement casing
2 liquid containers for oil or ice/water
3 different liquids
4 blackbody targets
5 Surface temperature sensor calibration sleeve
6 fixed point rooms

  • Water triple point room
    Gallium melting point chamber
  • Indium melting point chamber

7 Carrying Boxes
8 standard probes
9 Sensors/thermometers for options 1 and 2
10 RS232 communication or RS422 communication







Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!