Shanghai Senpu Technology Co., Ltd
Home>Products>SYNPEC P200 portable gas chromatograph
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
  • Address
    3rd Floor, Building 2, No. 118, Lane 133, Guangzhong Road, Minhang District, Shanghai
Contact Now
SYNPEC P200 portable gas chromatograph
Product Overview: Senpu P200 portable gas chromatograph with high integration; Easy and portable; Accurate measurement; Features such as easy operatio
Product details

Product Overview

SenpuP200Portable gas chromatograph with high integration; Easy and portable; Accurate measurement; Features such as easy operation. Compliant with national standards

HJ 1012-2018 《Environmental air and exhaust gasesTechnical requirements and detection methods for portable monitoring instruments for total hydrocarbons, methane, and non methane total hydrocarbons How to translate

HJ38-2017Determination of Total Hydrocarbons, Methane, and Non Methane Total Hydrocarbons in Waste Gas from Fixed Pollution Sources by Gas Chromatography


Working principle: According to the preprocessing method of Chinese national standards, the product first collects the sample gas to be tested and enters the portable gas chromatograph for quantitative analysis,

By switching the valve into the chromatographic column for separation, total hydrocarbons, methane, or characteristic factors sequentially reach the hydrogen flame ionization detector(FID)Detect and measure the concentrations of multiple components separately. Finally, the accurate value is calculated by the built-in processorTotal hydrocarbons, methane, non methane total hydrocarbons, and characteristic factor values.107 Product Features

Specialized for miniaturizationFIDDetector, meeting the requirements

Dynamic linear range, supporting automatic ignition, automatic judgment, and automatic shutdown protection

Integrate the traditional methane non methane total hydrocarbon online monitoring system into the same portable chassis, including a displayFID

Detector, switching valve, chromatographic column, electronic flow controller, battery, gas cylinder, etcNo need to connect to the mains or external gas lines on site, just one click to start up and measure the resultsTo ensure stable use of the instrument, the main unit and the heat tracing pipeline are powered by independent portable power sources

Fully automatic electronic flow modules are required for gas circuits such as carrier gas, hydrogen gas, and combustion assisting gas(EPC)Control. The precision of carrier gas pressure control has reached

0.001psiThe analyzer adopts a color touch screen, which can display the test concentration, test spectrum, and instrument parameter status. The analytical instrument can be connected to a mobile handheld terminal through a stable wireless connection, supporting functions on the handheld terminalApp120Perform control200Full process high-temperature heat tracing: from sampling toFID


The detector adopts a full process approach

to

High temperature heat tracing effectively solves the problem of condensation loss of samples in high-temperature and high humidity gas environments, providing more accurate test results

Main parameters

function

Analyze methane; Total hydrocarbons; Non methane total hydrocarbons; Benzene derivatives, etcV O C

s

size410 mm ( L ) x 280107 mm ( W ) x 235 mm ( H )

host weight

≤9K g

monitor

FID

Detector; satisfy

Dynamic linear range

Detection Principle

0GC - FIDChromatographic column separationFlow control accuracy

0.001 psi

analysis cycleNMHC:30smeasurement range

~5

0000ppm; AdjustableMinimum detection limit

0.03mg/m³

repetitiveness

RSD

0.5%accuracy

≤ Range ±

1 %6control panel

Portable high-definition industrial tablet

communicationGPSPositioning, wireless printing, data uploading, remote control200power supply

Built in power supply, with a battery life greater than

200hourprobe temperature

Heat tracing temperature

-15room temperature50

℃ adjustable

Valve box temperature


Max .

ambient temperature

℃ ~

1

working environment

10 - 90 % R H

1

2

Main configuration of the instrument

ITEM

1

3

CAT NO.

DESCRIPTION

1

4

QTY

P200Portable chassis for gas chromatograph host

1

5

detector

Hydrogen flame ionization detector

1

6

Sampling gunEquipped with a filtered heat tracing sampling gunsampling pump

KNF

1

7

Micro sampling pump

Flow path control

1

8

Fully electronic flow control

chromatographic column

1

9

/

Column box

1


Online inquiry