SYLVAC-P12D Ultra High Precision USB Interface Micrometer
Specially designed induction measurement probes for measurement
There are two versions: spring push and pneumatic push
Long life precision bearing guidance (tested up to 13 million times)
Resolution<0.0001 millimeters
• Repeatability 0.15 μ m
• Measurement range:+/- 2mm and+/-5mm
Compatible with D300SV2, D70i, and D400S units through an m-bus multiplexer
• High data speed
Operating temperature+10+80 ° C
IP65 protection
SYLVAC Inductive Probe P2i/P2iP/P5i/P5iP Technical Parameters
model | P2i | P2iP | P5i | P5iP |
Order number | 900.1101 | 900.1102 | 900.1111 | 900.1112 |
measuring range | ±2mm | ±2mm | ±5mm | ±5mm |
Pre itinerary | 0.15mm | 0.8mm | 0.15mm | 0.3mm |
maximum error | 1.35mm | 1.2mm | 0.85mm | 0.7mm |
resolving power | Digital : user selectable to <0.1μm | |||
Repeatability | 0.15μmm | 0.15μm | 0.15μm | 0.15μm |
Moving mass | 3.4g | 3.4g | 3.7g | 3.7g |
Accuracy of reading | 0.50% | 0.50% | 0.50% | 0.50% |
Main body diameter | 8h6(mm) | |||
Protection level | IP65 | |||
Temperature error | 0.01%FS/°C | |||
Standard | 0.7N | 0.7 / 0.4 bar - 2.6 / 1 bar N | 0.7N | 0.7 / 0.4 bar - 2.6 / 1 bar N |
Tolerance ± 20%, measuring probe in middle course, outgoing spindle.
Pneumatic probe
The pneumatic probe is fully retracted in the absence of air. In order to ensure continuous and reliable operation and maximize working life, the air supply should be clean and dry. 60% of maximum relative humidity, filtering particles larger than 2um. Air pressure: 0.4-0.8 bar, do not exceed 0.8 bar as the rubber sleeve may be damaged.
SYLVAC Inductive Probe P2i/P2iP/P5i/P5iP Dimensional Drawing
P2i
P2iP
P5i
P5iP
Application of SYLVAC-P12D Ultra High Precision USB Interface Micrometer:
SYLCOM software
D300S display unit
D62S display unit