1. LCD large screen (320 * 240) displaying both Chinese and English. Parameter settings are more convenient.
It can store 10 sets of chromatographic analysis data and can be accessed at any time.
2. It has a gas cutoff protection function to maximize the protection of the chromatographic column. Power outage bridge current protection
The device effectively protects rhenium tungsten wire.
3. A reliable anti-interference system that will not cause power fluctuations or momentary power outages
Affects the normal operation of the instrument.
4. Complete temperature overheating protection and platinum resistance open circuit and short circuit alarm functions to ensure temperature
The degree is not out of control.
5. Powerful self diagnostic function, allowing users to easily know the cause and location of problems.
6. The fully stable voltage and flow mode ensures precise and stable airflow. Powerful pneumatic control system, capable of
Expand to 21 routes.
7. High performance capillary dedicated vaporization chamber achieves true division/non division
Flow. This high-performance capillary device has obtained a national patent.
8. Can install 4 detectors simultaneously. Can be dual FID or dual TCD or dual ECD or dual
NPD operation. TCD is equipped with dual columns for supplementing qi, so it can be used for single column, double column, or capillary
Use of thin tube columns. Automatic ignition. Baseline zeroing, digital display. Rear door implementation
Fuzzy control.
9. Can simultaneously install and operate 3 sets of capillary columns and 3 detectors.
10. Innovative second chromatography column box module, which can flexibly achieve multi column chamber chromatography analysis work
Mode.
11. The detector can be used in series or parallel, and intelligent automatic conversion can be achieved through event control.
12. The instrument can be installed with a conversion furnace, thermal desorption instrument, automatic sampler, and automatic headspace injection
Various injection methods such as apparatus, disruptor, etc. Timed automatic sampling is available to achieve automatic sampling
Line analysis.
13. "One click Full Start" - that is, simultaneously starting the program to heat up, 2 external events
Multi channel workstation.
14. Full tail blowing system - that is, all detectors have their own independent tail blowing system
The system, including the thermal conductivity pool detector, also has 2 independent tail blowing systems that can be connected to small ones
Caliber capillary tube for thermal conductivity.
1. Temperature control range: room temperature+6 ℃~400 ℃ increments 1 ℃ accuracy ± 0.1 ℃
Program heating with 8th order heating rate ranging from 0.1 ℃ to -40 ℃/min
2.FID: D≤5×10-12G/s (n-hexadecane/isooctane)
Rn≤2×10-14A
Rd≤1.5×10-13A/30min
Linear range: ≥ ten
3. TCD: S ≥ 10000mv. ml/mg (benzene/toluene)
Rn≤0.02mv
Rd≤0.15mv/30min
Linear range: ≥ ten4
4.ECD:D≤3×10-14G/ml (γ -666/iso octane)
Rn≤0.04mv
Rd≤0.15mv/30min
Linear range: ≥ ten4
Radioactive source Ni63
5.FPD: D(p)≤1×10-12G/s (1605/ethanol)
D(s)≤1×10-12G/s (phenol/n-hexane)
Rn≤4×10-13A
Rd≤2×10-12A/30min
Linear range: ≥ ten3
6.NPD:D(p)≤5×10-13G/s (Malathion/Isooctane)
D(N)≤7×10-13G/s (azobenzene/isooctane)
Rn≤4×10-13A
Rd≤2×10-12A/30min
Linear range: ≥ ten5
7. Other:
Dimensions: 560 × 510 × 530mm
(Length x Width x Height)
Column installation span: 152.4mm
Instrument weight: 50Kg
Maximum power: 2500W
Voltage: 220 ± 10% 50Hz