Instrument Introduction:
SP-3887AA Zeeman Atomic Absorption Spectrometer
For the Zeeman single graphite furnace atomic absorption machine.
The outstanding performance of SP-3887 instrument includes:
Innovative in key technologies such as graphite furnace DC power supply lateral heating technology, automatic compensation of graphite furnace resistance change technology, lateral variable AC magnetic field Zeeman background correction technology, self-priming background correction technology, etc,
This product is a graphite furnace integrated machine that can effectively detect small signals from high background samples and directly detect high-temperature elements, with a wide range of application prospects.
Application areas and purposes:
The instrument is positioned in the mid-range upper price range and performance, with multiple different configurations to meet the needs of different users. The price ranges from 250000 yuan to 450000 yuan. Instruments can be widely used in fields such as food, medicine, environment, biology, agriculture, petrochemicals, construction, materials, geology, metallurgy, and scientific research. Given the excellent AC and DC Zeeman background and deuterium lamp background correction capabilities of the instrument, this background correction system can not only effectively correct traditional molecular absorption and particle scattering backgrounds, but also correct structural backgrounds and partial spectral line overlap interference. It is particularly suitable for the analysis and testing of trace elements in samples such as food, traditional Chinese medicine, seawater, blood, biological products, and high salt solutions. Especially for daily essential elements such as Cd, Pb, Cu, Zn, etc., the calibration performance can be comparable to imported Zeeman instruments, but the instrument price is only half of that of imported Zeeman instruments, which will be more suitable for China's national conditions.
Dual detector optical dual beam, total reflection achromatic optical system;
The super strong anti atomizer light radiation ability can effectively eliminate the strong light radiation of graphite tubes at high temperatures.
Intelligent stray light measurement and correction technology, Effectively improved the optical accuracy and linear range of the atomic absorption spectrometer, and enhanced its background correction capability.
Optical noise reduction technology, combined with optical component ultraviolet enhancement technology, effectively improves the optical performance and linear range of the instrument.
Gap control lamp technology enables ordinary domestic hollow cathode lamps to be used for self-priming background correction, extending the service life of hollow cathode lamps by more than ten times and reducing usage costs.Horizontal Zeeman magnetic field longitudinal heating graphite furnace
Adopting an eight lamp rack system with light source emission direction parallel to the center of gravity, automatic lamp replacement.
Simultaneously equipped with three background fastening methods: deuterium lamp, self-priming, and Zeeman;
Wavelength range: 180nm-900nm Automatic peak searching
Spectral bandwidth: six automatic switching modes: 0.1nm, 0.2nm, 0.4nm, 0.7nm, 1.4nm, 2.0nm
The position of the graphite furnace atomizer can be adjusted horizontally, vertically, and in reverse without any adjustment;
Graphite Furnace Visualization System
Measurement methods: graphite furnace method, hydride method;
Concentration calculation methods: standard curve method (a total of 6 linear and nonlinear fitting methods), standard addition method, internal standard method.
Graphite furnace Zeeman background correction:>100x (1Abs);
Graphite furnace automatic sampling technology, with sample cup positions ≥ 80.