SDW-163Portable oil particle counterintroduce
SDW-163 portable oil particle counterCompliant with DL/T 432, GB/T 20082, ISO 4406, NAS 1638, GB/T 18854, ISO 11171, GJB 420B, GJB 380.4A, NAS 1638, it is used to measure the particle pollution level of phosphate ester fire-resistant oil, turbine oil, transformer oil, and various other auxiliary engine oils. Especially suitable for oil samples with severe emulsification or high viscosity. The determination method is to vacuum filter the oil sample, so that the particles in the oil sample are evenly distributed on the microporous filter membrane. Under the transmitted light of the oil contamination comparison microscope, it is compared with the oil contamination grading standard template to determine the particle contamination level of the oil sample.
Functional features:Easy to operate, suitable for on-site testing
Vacuum suction filtration device, which can accelerate the filtration speed
Standard template for comparing pollution levels
Technical parameters:
Microscope 100x (with illumination)
Detecting particle size ≥ 5 μ m
Level range NAS level 4-12
Dimensions: 410mm × 310mm × 230mm
Weight 6.4kg