R500 ProSurface contaminationinstrument
sketch
R500 ProThe surface pollution meter is excellent due to itsshapeDesigned, obtainedAMTInnovative Design Award.Koley Company'sNewly developedR500 ProSurface pollution meter, usingCOLIYThe company's patented designnew typeultrathinDouble flash detector,Not using commonly used large volume photomultiplier tubes, resulting in a detector thickness of only16mm,The effective detection area is15cm2No need for inflation, greatly reducing the cost of useImproved usability.Optional cesium iodide gamma ray sensor and ultra-thin sensorDouble flash detectorPlaced on the same plane, so thatR500 ProSurface pollution meterCan simultaneously detect alpha, beta, and gamma rays. ultrathinDouble flash detectorusedThe thickness of the plastic scintillator is0.5mmIt greatly reduces the interference of gamma rays. The thickness of the aluminum film on the incident window is less than3μmYes, that's rightαThe absorption of radiation is only1MeV.The host adopts a color screen,Can simultaneously detect and display alpha, beta, and gamma ray values.
R500 ProSurface pollution meterCan quickly and reliably screen for radiation pollution,Directly detecting the surface of an objectα、βLevel of radioactive surface contaminationandγ-rayHigh detection efficiency and fast response speed,R500 ProSurface pollution meterFully meet the needs of usersHighly sensitive and multifunctional Requirements for ease of operation, etc.R500 ProSurface pollution meterErgonomically designed, the grip is easy to grip and the palm will not be contaminated. Equipped with correction factor function, customers can adjust the correction parameters themselves; The patented algorithm provides fast response and significantly improves sensitivity while maintaining stable background values. The anti saturation circuit can avoid interference from over range and electromagnetic waves, and has a safety calibration function that can prevent radiation exposure by calibration personnel.
application
R500 ProSurface pollution meter, applicable toEnvironmental protection, military, customs, health, nuclear industry, civil aviation, pharmaceuticals, nuclear power plants and other systems.
useMonitoring of radioactive environments, agricultural markets, workplaces, laboratories, hospitals, isotope production plants, nuclear material transportation, storage, and inspection sites, monitoring of human radioactive contamination levels, monitoring of goods radioactive contamination levels, food, and operationsMonitoring of radioactive contamination on surfaces such as surfaces, floors, walls, hands, and clothing.
Product characteristics
1. Can simultaneously detect alpha, beta, and gamma rays
2. Can display simultaneouslyα. Beta and gamma radiation values
3. Patent Designnew typeultrathinDouble flash detector
4. The thickness of the detector is only16mm,No need to inflate
5. Effective detection area15cm²
6. Probe adoptsZnSadd0.5mmThin plastic scintillator, no need to inflate
7. Aluminum film covering the entrance window, with a thickness less than3μm
8. Detecting background:αthoroughfare≤ 0.1cps;βthoroughfare≤ 15cps
9. Detection efficiency:α≥30%(239Pu);β≥25%
10. Optional cesium iodide gamma ray sensor
11. Can be connected through the upper computer softwarePCImplement data storage, transmission, and analysis, etc
Technical Parameter
Measure the type of radiation |
α、β、γandΧRadiation (withprotectSliding cover) |
α、βsurfacepollutionDetecting background |
αthoroughfare≤ 0.1cps;βthoroughfare≤ 10cps |
α、βsurfacepollutiondetection efficiency |
α≥30%(239Pu);β≥25% |
α、βSurface contaminationDetection area |
15cm2 |
α、βSurface contaminationDetection sensor |
Ultra thin dual flash detector(ZnScoating+Thin plastic scintillator) |
α、βSurface contaminationDetection range |
0-105cps 0105 Bq/cm2 |
γandΧraysensor |
Cesium iodide crystal sensor |
γandΧRadiation measurement range |
0-105cps 0.01µSv/h-5000µSv/h |
γandΧX-ray sensitivity |
1µSv/h≥20cps(Cs137) |
γandΧRadiation energy response |
80keV -2MeV |
display |
black and white128x64LCDdisplay |
Alarm |
Sound and light alarm, alarm value can be freely set within the range |
accuracy |
±15% |
communication interface |
USB |
Magnetic field influence |
Magnetic field interference is noteffectmeasure |
operation temperature |
-20℃to+50℃ |
storage temperature |
-20℃to+60℃ |
Working humidity |
80%R.H(Non condensing) |
Working atmospheric pressure |
75kPa-110kPa |
weight |
450can |
size |
L300mm;W90mm;H40mm |
power supply |
3joint5namealkalinebatteryOr rechargeable battery |
Certificate |
CE,ISO9001 |
COLIYPatented ultra-thin dual flash sensor
After damage, it can be replaced as a whole to reduce user costs
1. Slide cover
Open the slide cover to detectα、β、γandΧray
Close the sliding cover, the sensor receives dual protection and can detectγandΧray
2.Telescopic rod
Telescopic pole, longest1.53Meter, can measure nuclear radiation from a distance, avoiding human injury
The front end angle of the telescopic rod can be adjusted freely
Real time remote transmission of numerical values to the computer for display and analysis
Upper computer software