Portable COD ammonia nitrogen total phosphorus analyzer
YCNP-3P portable COD ammonia nitrogen total phosphorus analyzer, using the following standards: HJ/T399-2007 ,HJ535-2009,GB11893。 According to Lambert
Product details
YCNP-3P PortableCOD ammonia nitrogen total phosphorus analyzerTECHNICAL INDEX
- Measurement range: COD: 5~10000mg/L Ammonia nitrogen: 0.05~50mg/L Total phosphorus: 0.02~20mg/L
- Measurement error: COD, total phosphorus ≤ ± 5%, ammonia nitrogen: ≤ ±3%
- Detection limit: COD: 5mg/L, ammonia nitrogen: 0.05mg/L, total phosphorus: 0.02mg/L
- Repeatability: ≤ 3%
- Optical stability: ≤ ± 0.001A/20 minutes
- Anti chlorine interference: high range ≤ 2000mg/L, low range ≤ 1000mg/L
- Dissolve: COD: 165 ℃ (15 minutes)
- Custom settings: Temperature (room temperature -200 ℃)
- Temperature control error: ± ≤ 1.0 ℃
- Colorimetric method: Colorimetric tube
- Batch processing capacity: 6 samples
- Size: Host: 200mm * 150mm * 90mm Dissolver: 210mm * 150mm * 90mm
- Power supply: 2800mA rechargeable power supply for the host, optional car power cord for the digestion instrument
- Weight: Host < 1.5kg Dissolver: 2.0kg
- Environmental temperature: 5-40 ℃ Environmental humidity: ≤ 85% No condensation
YCNP-3P typePortable COD ammonia nitrogen total phosphorus analyzerInstrument characteristics
- LCD screen interface design, fully Chinese operation, direct data reading, simple operation.
- Imported cold light source, no preheating required, energy-saving, environmentally friendly, fast response, with a lifespan of up to 100000 hours and strong optical stability.
- Built in standard curve, no need for user calibration and calibration.
- Enable guided measurement with one click operation, simple and fast
- Adopting a digestion colorimetric system, reducing manual operation, the measurement is simple, fast, and safe.
- Customize more than 40 curves and store 200 complete detection data.
- Intelligent PID temperature control technology and dual anti overheating protection system ensure safe, uniform, and fast heating.
- Standard 6-hole intelligent digestion instrument, which can achieve one click digestion without setting.
- Custom mode, temperature (room temperature -200 ℃) can be set arbitrarily.
- Equipped with power-off protection function to prevent data loss.
- One click factory reset to prevent loss or accidental deletion of standard curves.
- The equipment is not damaged by human factors, with a three-year warranty and software upgrades
Online inquiry