Photochemical Reactor CT-GHX50 Safety:
At all stages of operation, maintenance, and instrumentation, the following basic safety measures must be followed. When using the instrument, follow the instructions to operate it,Unauthorized use can cause the normal operation of the instrument and even damage it.
ATo avoid electric shock accidents, the input power cord of the instrument must be grounded. This instrument uses a three core grounding plug, which has a grounding pin. If the plug cannot be inserted into the socket, an electrician should install the correct socket to prevent the instrument from losing its grounding protection function.
BPay attention to the use of power supply:
Before connecting the AC power supply, ensure that the voltage is consistent with the voltage required by the instrument (allowable ±)10%Ensure that the rated load of the power socket is not less than the requirements of the instrument.
CPay attention to using power cords:
This instrument usually uses the power cord that comes with it. If the power cord is damaged, it must be replaced and cannot be repaired. When replacing, a power cord of the same type and specification must be used instead. No objects are allowed to be placed on the power cord when using this instrument. Do not place the power cord in areas where personnel are walking.
DPay attention to the placement of instruments:
The photochemical reactor CT-GHX50 should be placed in a cool, ventilated, dry, and dust-proof location. For better heat dissipation, the instrument should be kept at an effective distance from other objects in a well ventilated area(N﹥30cm).
Photochemical Reactor CT-GHX50Scope of application:
The photochemical reaction instrument CT-GHX50, developed in collaboration with outstanding domestic laboratory practices, is a new generation of photochemical reaction device. The photochemical reaction instrument CT-GHX50 is mainly suitable for studying whether gas phase, liquid phase, solid phase, and flow systems are loaded under simulated ultraviolet light, simulated sunlight, simulated visible light, and special simulated light irradiationTiO2Photochemical reactions under conditions such as photocatalysts are used in research fields such as chemical synthesis, environmental protection, and life sciences. The system has unique advantages such as reasonable technology, simple structure, touch sensitive buttons, convenient operation, stable operation, protection of the human body, free combination, and flexible customization.Product features of photochemical reactor: 1. The main control power supply of the photochemical reactor CT-GHX50 is flexible in controlling the illumination time, suitable for timed operations and experimental use; twoA professional and stable simulated light source with a space saving volume design, especially suitable for laboratories with limited space;3.There are multiple test tube magnetic stirrer reactors in the supporting device, which can achieve partial tube inflation at the same time. The practical application value and performance of multiple test tube magnetic stirrer reactors are * *;4.Equipped with a multi port magnetic stirring reactor function, it can have strong magnetic stirring, inflation, deflation, sealing, measurement and other functions during the reaction process5.Equipped with a solid reaction device, it can perform photocatalytic reactions on solid substances to improve catalytic speed;6.Equipped with cooling water system and imported compressorFluorine free operation ensures long-term stable operation of the light source, suitable for continuous operation experiments. The low-temperature cooling water supply device is equipped with a silent external circulation pump to provide cooling water circulation and pressurization, while saving water waste; 7 Photochemical Reactor CT-GHX50Special mixer is a new type of specialized strong magnetic stirring device that can evenly and uniformly irradiate the sample, achieve synchronous stirring of multiple test tubes, and have multiple functions. The stirring magnet in each test tube can be infinitely adjusted in speed.
8. The main components of this machine are selected with excellent materials, the motor is a high-power DC motor, the stirring force is huge, the noise is low, and the magnetic steel is selected as the current strongest magnetic "neodymium, iron, and jumping" water magnet for the rotor, ensuring sufficient sharpness and torque
Advantages and characteristics of the photochemical reactor CT-GHX50
1. There are mainly various light source devices, glass reaction vessels, all quartz light source protection cold traps, and dedicated parallel devices. It consists of a rotating reactor, a main reaction protection box, and an inflation control device.
2. It can be used to determine reaction kinetics and constants, provide analysis of reaction products and quantum rates, and for the development of green energy.
3. All quartz reaction vessels are divided into:ATheBType, meet different customer needs, (also accept customized non-standard reactors)
4. The high and low temperature cycler provides light source protection, cold trap cooling or constant temperature reaction, and flexibly matches different power light sources to improve the efficiency of the high and low temperature constant temperature cycling device (-20degree-100Degree)
5. The machine is equipped with a UV filter device to ensure the optimization of the light source.
6. The configuration of this model can be recommended based on the actual usage situation of the customer, or a customized photochemical reaction device that meets the actual requirements of the customer can be provided
Configuration List
1、 Specification and model: CT-GHX50
2、 Product Name: Photochemical Reactor
3、 Configuration standards:1.The main reaction part includes a reaction chamber, mercury lamp electrical controller, xenon lamp electrical controller, glass reaction vessel, light source, and magnetic stirrer;
2.The mercury lamp electrical controller is used for control300WThe500WWorking conditions of medium pressure mercury lamps; optional100WThe200WThe1000WThe1500WWaiting for specifications)
3.Xenon lamp electrical controller for control500WThe1000WWorking conditions of xenon lamp; optional250WThe350WThe800WThe1500WWaiting for specifications)
4.All quartz glass reaction vessels can be selected separately20MLThe50MLThe60ML、100ML、300ML、500ML、1000ML 等
5.All quartz glass cold trap (extended cooling cycle at the bottom of the cooling water)
6.Configure the photochemical reactor CT-GHX20Specialized parallel reaction magnetic stirrer;
7.The cooling water circulation device can achieve-20degree-100Temperature control of * * degree.
photoelectric parameters
The photoelectric parameters of the high-pressure mercury lamp are as follows:
|
Power W |
Starting current A |
Working current A |
Working voltage v |
Effective arc length MM |
wiring method |
|
100 |
1.5 |
0.4 |
145.8 |
75±10 |
Single ended lead out |
|
300 |
5.7 |
3.4 |
138.1 |
125±10 |
Single ended lead out |
|
500 |
7.1 |
5.4 |
140 |
170±10 |
Single ended lead out |
|
1000 |
17.3 |
10.4 |
143.2 |
250±10 |
Single ended lead out |
The relative intensity of high-pressure mercury lamps is as follows:
|
wavelengthnm |
250 |
313 |
365 |
400 |
510 |
620 |
720 |
|
Current strength% |
20 |
85 |
100 |
30 |
20 |
40 |
80 |
The photoelectric parameters of xenon lamps are as follows:
Dimensions and photoelectric parameters of xenon lamps
|
Power W |
Starting current A |
Working current A |
Working voltage V |
Effective arc length MM |
wiring method |
|
350 |
5.5 |
5.0 |
73.1 |
190±10 |
Single line method |
|
500 |
8.3 |
6.4 |
75.6 |
190±10 |
Single line method |
|
800 |
11.4 |
10.0 |
83.5 |
220±10 |
Single line method |
|
1000 |
12.4 |
12.1 |
87.6 |
220±10 |
Single line method |
The photoelectric parameters of mercury xenon lamp are as follows:
Dimensions and photoelectric parameters of mercury xenon lamps
|
Power W |
Starting current A |
Working current A |
Working voltage V |
Effective arc length MM |
wiring method |
|
300 |
6.3 |
4.1 |
220±20 |
175±10 |
Single line method |
|
500 |
9.7 |
6.2 |
220±20 |
190±10 |
Single line method |
|
1000 |
19.1 |
11.7 |
220±20 |
200±10 |
Single line method |
The photoelectric parameters of metal halide lamps are as follows:
Dimensions and photoelectric parameters of metal halide lamps
|
Power W |
Starting current A |
Working current A |
Working voltage V |
Effective arc length MM |
wiring method |
|
250 |
4.8 |
3.0 |
80.7 |
35±10 |
Single line method |
|
400 |
10.7 |
5.4 |
80.2 |
35±10 |
Single line method |
|
800 |
26.73 |
10.3 |
90.9 |
45±10 |
Single line method |
1.Mercury lamp, xenon lamp turned on5After minutes, the light will be stable. After turning off the lights, you need to wait5It can only be restarted in minutes;
2.The emitting part of mercury and xenon lamps is a quartz tube, and direct contact with hands can cause pollution and affect the luminous intensity. Quartz
Clean the outside of the vessel with a small amount of alcohol dipped in a clean fat cotton, and clean the inside with detergent;
3.Multiple tube magnetic stirrer reaction tubes are selected separately30ml,50ml,60ml、100ml、300ml、500ml、1000mlwait.
4.The main box is the place where the series of chemical reactor lighting devices are placed, and it has the following characteristics: box body
The size is440mm*440mm*770nm 约26KG
5.Ambient temperature0℃-35℃Relative humidity≤80%.
6.warranty period:a year