- Product Name: Oxygen permeability tester - Electricity method
- Product Code: OTR-E-11-A
OTR-E-11-A
Function
Oxygen permeability performance test.
Films, sheets, boards, containers made of materials such as plastic, fabric, leather, metal, etc.
Working principle
Coulomb electric quantity method.
Oxygen with a certain humidity or dryness flows on one side of the material, while high-purity nitrogen (carrier gas) flows at a fixed flow rate on the other side; The difference in oxygen concentration on both sides of the sample drives oxygen to permeate from the oxygen side to the nitrogen side of the sample; The permeated oxygen is carried by nitrogen (carrier gas) to the coulomb meter micro oxygen sensor; Sensors measure the oxygen content of the carrier gas and calculate parameters such as the oxygen permeation rate of the sample.

Compliant with standards
YBB 00082003、GB/T 19789、ASTM D3985、ASTM F2622、ASTM F1307、ASTM F1927、ISO 15105-2、JIS K7126-B
Product Features
- Original imported oxygen sensor, ppb level.
- Humidity control: 35%~95% RH, 100% RH, fully automatic, no fog.
- Temperature control: Semiconductor cold and hot bidirectional control, high precision, stable and reliable.
- Super environmental adaptability: Indoor environment, temperature range of 10 ℃ -30 ℃, no humidity requirements, low usage cost.
- Fully automatic: no mechanical knob; One click start, fully intelligent throughout the process; Power off data storage.
- Sample anti leakage sealing installation technology; Self detection of startup faults to avoid testing under fault conditions.
- Software: graphical, full process, and all element monitoring; Multiple report formats.
- Optional: GMP "Computerized System" functional module.
TECHNICAL INDEX
|
Name |
Parameters |
name |
parameter |
|
Humidity range |
0% RH,35%~95%RH,100%RH |
Moisture control error |
±1%RH |
|
temperature range |
15℃~50℃ |
Temperature control error |
±0.1℃ |
|
specimen thickness |
<3 mm |
Mpa |
≥0.2 MPa |
|
Carrier gas type |
Nitrogen with a purity of 99.999% |
carrier gas flow |
0~100 cm3/min |
|
interface size |
1/8 inch metal tube |
Host size |
720(L)×415(W)×400(H) |
|
Power source |
AC 220V 50Hz |
Power supply power |
1200W |
|
Model differentiation |
Model number |
||
|
OTR-E-11-A |
|||
|
Sample quantity |
1 |
||
|
Sample area (cm)2) |
50 |
||
|
Test scope (membrane) (cm3/m2.24h) |
0.01~14,000 |
||
|
Resolution (membrane) (cm3/m2.24h) |
0.01 |
||
|
Test scope (container) (cm3/pkg.24h) |
0.00005~70 |
||
|
Resolution (container) (cm3/m2.24h) |
0.00005 |
||
|
Net weight (kg) |
52 |
||
system configurationHost, testing computer, professional testing software, Agilent oxygen trap, sampler, gas cylinder pressure reducing valve, sealing grease.
Select accessoriesContainer testing component, container temperature control testing component.
Self provided partsHigh purity nitrogen, high-purity oxygen, distilled water.