The Cermet II I.S. is an all-in-one humidity meter that is easy to install and used to measure the moisture content in flammable or common gases. Its special design is suitable for hazardous environments. It includes an inherently safe transmitter Transmet I.S. and a safety application control module Cermet II I.S. Under the stainless steel casing, the sealing protection level of Transmet I.S. reaches IP66 (NEMA 4X).
The Cermet II I.S. control module is a 19 'rack mounted component with display, voltage regulator module, and safety barrier, suitable for safe areas. This humidity meter provides all necessary interfaces for users, including multi unit display, alarm, analog and digital output, etc. The cable between the control module and the transmitter through the safety barrier can be up to 500m in length.
highlight
- ATEX, FM, and CSA certifications
- Pressure sensor input
- Measurement accuracy ± 1 ° C dew point
- User selectable measurement unit
- Up to 4 channels of alarms
- Wide range, calibration range -100~+20 ° C dew point
- Analog and digital outputs
|
 |
technical indicators
|
| Humidity range |
-Dew point at 100~+20 ° C |
| precision |
± 1 ° C+20~-60 ° C dew point ± 2 ° C -60~-100 ° C dew point |
| other units |
Digital processing and linearisation with pressure compensation from measured dew point with pressure compensation to any other preferred unit-lb/MMSCF and mg/Nm³ (for natural gas), ppmVwith active or passive pressure compensation |
| Pressure input |
Fixed value (user programmed) or dynamic measurement with an optional 4-20 mA pressure transmitter |
| resolution |
0.1 °C dp, 0.1 °F dp, 0.01 ppmV, 0.1 mg/m³, 0.01 lb/MMSCF |
| analog output |
4-20 mA with isolation (maximum load 600 W) |
| Sensor calibration |
Tracing back to national humidity standards in the UK (NPL) and the US (NIST) |
| power supply |
Universal 85~265 V AC; 90 ~ 370 V DC |
| call the police |
Two user selection settings |
| Installation panel |
10 'aluminum, 3U height 132 x 200mm depth with gap |
| Additional voltage regulator source |
Used for Transmet I.S. 24 V DC |
| safety barrier |
2-way isolated safety barrier |
| Control module options |
|
| power supply |
Low voltage type 18~36 V AC; 9 ~ 60 V DC |
| relay |
1 or 2 5A relays (up to 4), normally open type, user adjustable: full range+all unit selection+sensor fault alarm function |
| output |
Simulation: 0-10V user set corresponding range Number: RS485: Provides 2-channel data transmission to users+all setup work |
| TRANSMET I.S. transmitter |
|
| Sensor thread |
5/8 'UNF Parallel |
| shell |
stainless steel |
| Sealing protection level |
IP66 (NEMA 4X) |
| power supply |
On the Cermet II I.S. control module |
| Operating Temperature |
-20 ~ +60°C |
| Storage temperature |
-40 ~ +70°C |
| work pressure |
Up to 30 MPa (300 barg) |
| traffic |
0-10 Nl/min |
| gas flow rate |
20m/sec (最大) |
| Sensor connection cable options: |
| On site cable indicators for polyethylene insulated instrument cables refer to BS5308 Part 1 |
|
| Polyethylene insulation, metal wire isolation shielding, PVC blue jacket. Other similar materials are available for selection |
| Maximum length of 500m (depending on the technical specifications of the cable) |
| safety |
| The Cermet II I.S. humidity meter complies with the standard BS5345: Part 1:1989 and other relevant European safety regulations for the operation of sensors in hazardous environments and the sampling system for analyzing gases. The Cermet II I.S. control module must be placed in a non hazardous location (or in an optional EExd explosion-proof box). Transmet I.S. is ATEX certified and can be used in hazardous environments II 1G EEx ia IIC T4 after using a pair of isolated or Zener safety barriers on signal and power connections. Transmet I.S. also meets FM certification (USA) for CL1, Div1, GpA, B, C&D, T4, as well as CSA (Canada) certification for Ex ia (Class 1, Groups A, B, C&D). |
|
|
|
|
|
|