QE Pro is a highly sensitive spectrometer suitable for low light applications such as fluorescence measurement, DNA sequencing, and Raman analysis. The thin back illuminated CCD detector of QE Pro has high quantum efficiency, and its robust design gives it high signal-to-noise ratio performance and stability. The measurement of dark noise can be optimized by selecting an internal shutter.
Product Features
90% extremely high quantum efficiency and extremely low etalon effect
● Signal to noise ratio (>1000:1) and improved processing speed
TEC cooling (up to -15 ° C) allows the detector to have lower noise and is more conducive to low light detection
And the integration time can be continuously adjusted from 8ms to 15 minutes, and the spectrum will not be distorted or deformed
Provide a grating with 900 lines/mm (# H36), with a shining wavelength of 500nm, and good spectral response for 532nm and 785nm laser Raman spectroscopy
● Thermal stability
● Internal shutter (optional)
application
● Biotechnology applications ● Medical diagnosis ● Metallurgical analysis ● Protein and nucleic acid analysis ● Polymer analysis ● Chemistry
specifications
Engineering specifications |
QE Pro (user configured) |
PHYSICAL |
|
Size: (Length)xwidexHeight in millimeters and inches |
182Millimeter(7.17Inches)x 110Millimeter(4.33Inches)x 47Millimeter(1.85Inches) |
Weight (kilograms and pounds): |
a spectrometer:1.15Kilogram(2.6Pounds); Power Supply:0.45Kilogram(1Pound) |
DETECTOR |
|
Type: |
Hamamatsu S7031-1006 scientific grade, back-thinned, TE Cooled, CCD array |
Range: |
185-1100 nm (dependent on grating selection) |
Quantum efficiency: |
90% (peak) |
SPECTROSCOPIC |
|
Wavelength range: |
185-1100 nm, grating dependent |
Integral time: |
8 ms to 60 minutes |
Dynamic range: |
~85,000:1 |
Signal to Noise Ratio: |
System: 1000:1 (single acquisition) |
Grating: |
Multiple gratings can be selected(H1–H14),HC1raster |
Slit: |
5, 10, 25, 50, 100 or 200 μm wide slits (or SMA/FC bulkhead with no slit) |
Optical resolution: |
0.14 – 7.7 nm (depends on grating and size of entrance aperture) |
Scattered light: |
<0.08% at 600 nm; 0.4% at 435 nm |
Onboard cache: |
15,000 spectra |
Fiber optic connector: |
SMA 905 and Ocean Optics FC |
Internal shutter (optional): |
Actuation time: 11 ms |
ELECTRONICS |
|
Power requirements: |
Power supply voltage:4.5–5.5bend over |
Strobe function: |
Continuous and single flashes |
Interface: |
USB 2.0,480Mbps(USB 1.1Compatibility);RS-232(5Line) |
working conditions |
|
Working temperature: |
0 °C – 50 °C |
Humidity: |
less than or equal90%(No condensation) |