YS-12 standard COD digester
1. The new type of glass burr reflux tube and heat dissipation system ensures the parallelism and accuracy of sample digestion results due to its excellent digestion reflux effect.
2. Digital display of microcrystalline panel temperature. Digital display of digestion time: temperature and timing can be set at any time.
After digestion is complete,National standard COD standard digesterAutomatically stop heating, the instrument fan continues to work for 50 minutes to assist in sample cooling, and can be left unattended.
4. Energy and water conservation, safety and reliability have improved the work efficiency of operators.
IItechnologyparameter
Model number | YS-12 type |
Compliant with national standards | GB11914-89, Determination of Chemical Oxygen Demand for Water Quality, Potassium Dichromate Method. |
measuring range | Water samples with concentrations between 10-1000mg/L and greater than 1000mg/L |
Measure the dissolution time after dilution | 10 minutes to 2 hours |
measurement error | Potassium hydrogen phthalate standard solution (500mg/L) |
Relative standard deviation not exceeding 5.0%, industrial organic wastewater (500mg/L) | |
Relative standard deviation and deviation not exceeding 7.0% | |
ambient temperature | ≤35℃ |
Number of dissolved samples | 12 (using 250mL conical bottles with 24 # grinding mouth) |
Power | 1500W(AC 220V,50HZ) |
Cooling method | Air cooling, water cooling |
Leakage protection | Yes, there is |
3、 Delivery List
One digestion analyzer host, 12 condensers, 12 250ml conical flasks, and 2 fuses