Salt water spray testing machine is designed for surface treatment of various materials, including coating, electroplating, organic and inorganic film, anodic treatment, anti rust oil and other anti-corrosion treatments, and then test their corrosion resistance to determine the quality of products.
Nanjing Salt Spray Corrosion Test Chamber Manufacturer Price Promotion Low Price
The test case body is made of light gray impact resistant PVC board with a thickness of 5mm and a durable temperature of 85 ℃. Surrounding the edges with a three-dimensional reinforcement thickness of 10mm, the structure is sturdy
The sealing of the laboratory cover is made of PVC impact resistant board with a thickness of 5mm, which does not distort or deform during high temperature use. The inner and outer layers of the edge are thickened to prevent distortion and deformation caused by long-term high temperature. Thickness 15mm
C hidden integrated reagent replenishment bottle, easy to clean and operate
The saturated air tank is made of SUS # 304 stainless steel, which is resistant to high pressure and has excellent insulation effect
The sample placement table is made of 8mm heat-resistant ABS board with strong load-bearing capacity. The test sample holder adopts a flat indexing type, which can adjust the angle arbitrarily. The fog drops evenly on all four sides, and the fog is completely consistent. The test results are accurate, and the number of test samples placed is large. (Can be customized according to customer product requirements)
2 spray system
A adopts the Bernoulli principle to absorb saltwater and then atomize it uniformly, without any obstruction or crystallization, ensuring the standard of continuous testing
B Nozzle: made of special glass nozzle, which can adjust the size of spray volume and spray angle
Nanjing Salt Spray Corrosion Test Chamber Manufacturer Price Promotion Low Price
When the water level is low, the power is automatically cut off and the safety warning light device lights up to display
When the temperature exceeds 2 degrees Celsius, the heater power supply will be automatically cut off, and the safety warning light device will light up and display
When the water level of the test drug (saline) is low, the safety warning light device lights up and displays an alarm prompt
4. Leakage protection function to prevent personal injury and instrument failure caused by circuit leakage or short circuit

Operate and set according to JIS, ASTM, CNS, GB standard parameters
A Salt water spray test: NSS, ACSS
1 Laboratory:...... 35 ℃± 1 ℃ Volatility: ≤± 0.5 ℃ Uniformity: ≤ 2 ℃
2 saturated air tanks:...... 47 ℃± 1 ℃
B Corrosion resistance test: CASS
1 Laboratory: 50 ℃± 1 ℃ Volatility: ≤± 0.5 ℃ Uniformity: ≤ 2 ℃
2 saturated air tanks:...... 63 ℃± 1 ℃