Shanghai Minghe Instrument Co., Ltd
Home>Products>NT6108-AB200 type alpha and beta surface contamination detector
NT6108-AB200 type alpha and beta surface contamination detector
Product Introduction: The NT6108-AB200 surface contamination meter uses a ZnS coated plastic scintillator probe designed to simultaneously measure sur
Product details


Product Introduction

NT6108-AB200 type The surface contamination meter adoptsThe plastic scintillator probe coated with ZnS is designed for simultaneous measurement of surface contamination on α, β, and γ surfaces, and can be used for surface contamination detection on workbenches, floors, walls, clothing, skin, and other surfaces in radioactive workplaces or laboratories. Widely used in environmental protection, public security, military, customs, health, nuclear industry, nuclear power plants, universities, enterprises and other units and departments.

Product features

New plastic scintillator detector, no need to inflate

Can measure simultaneouslyAlpha and beta/gamma, without the need for an external probe

Monitoring and compensation of β/γ background, with continuously adjustable alarm thresholds

Adjustable measurement time

160*160Large area dot matrixLED display, capable of automatic sensing in dark environmentsAutomatically turn on the background light

Large detection area:200cm2

User friendly interface, displaying unitsCPS, Bq or Bq/cm2

Technical Specifications

Detector type: plastic scintillator+ZnS(Ag), The detection window adopts aluminum containing polyester film foil;

Detector size:200cm2

Alarm method: Sound and light alarm

UnitCPS, Bq or Bq/cm2

Background deduction: Deduction and non deduction of background are optional, and the measurement time of background can be set

Weight:750g (including battery);

Measurement range:0.01~9999kCPS ,0.01~9999kBq,0.01~9999k Bq/cm2

Background Count:α: ≤ 0.1 cps ,β/γ :≤ 10cps

Surface emissivity response:α ≥ 0.35 (for 241Am), β ≥ 0.25 (for 204TI);

Detection limit:9999KCPS;

The alarm threshold can be adjusted across the entire range;

Relative inherent error: not exceeding±20%;

Measurement repeatability: not exceeding±10%;

Can measure simultaneouslyα. β/γ, α or β/γ can be measured separately;

Power supply: Rechargeable lithium battery can work continuously48hour

Shell: Ergonomic plastic shell

Instrument size:320×150×120mm(L(含手柄)×W×H);

Usage environment

The working temperature is-20℃~+50℃

The working humidity is≤95%RH (30℃, non condensing)


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!