Testing method for residual ethylene oxide:
1.Principle of residual detection of ethylene oxide
At a certain temperature, extract epoxyethane from the sample with water and determine the epoxyethane content using headspace gas chromatography.
2.Gas chromatograph conditions for residual ethylene oxide detection
Hydrogen flame analyzer:Sensitivity not less than2X 1 0-11g/s(Benzene, carbon disulfide)(CS2)[]
chromatographic column:The chromatographic column used should be able to completely separate impurities and ethylene oxide in the sample, and have a certain degree of water resistance. Chromatographic column optional
| Column length | inner diameter | carrier | column temperature |
| 30m | 0.53mm | 30QC5/AC10-1.0 | 40Celsius |
| 1m-2m | 2mm-3mm | GDX-407 80eye~100eye | about130Celsius |
| P.-PA Q- 80eye~100eye | about120Celsius |
Temperature of various parts of the instrument:
a)Gasification room200centigrade;
b)Testing Room250Celsius
airflow rate:
a)nitrogen15m L/min-30m l-/min;
b)hydrogen30m L/min;
c)air300 mL/min
3.Preparation of standard storage solution for residual ethylene oxide testing
Take external dry50mCapacity bottle, with additional capacity30m 1Water, add bottle stopper, weigh accurately.0.1 m gInject approximately with a syringe0. 6 mLEthylene oxide, without a bottle stopper, gently shake well, cover the bottle stopper, weigh, and the difference between the two weighings is the weight of ethylene oxide contained in the solution. Add water to the mark and dilute the solution to1 X 10-' g/LAs a standard storage solution.
4.Preparation of residual ethylene oxide test
4.1The preparation of test samples should be carried out immediately after sampling, otherwise the test samples should be sealed in a metal container sealed with polytetrafluoroethylene for storage.
4.2Cut the sample into5 mmLong fragments, take2. 0 gPut it into the extraction container and add10 mLWater, Top Space40 mLThe pressure inside the container is at atmospheric pressure and is maintained in a constant temperature water bath60 Cscholar1'CMiddle placement20 min..
5.Testing steps for residual ethylene oxide
5 .1Prepare with reserve solution1X10-3g/L—1X102g/LStandard solutions with six series of concentrations. Each take10mLpress4 .2square
Legal processing.
Quickly take samples from the balanced standard and sample in sequence using a glass syringe1mLUpper gas, injected into the sampler.Record the peak height of ethylene oxide(Or area).
note1:Try to have one person operate in an analysis and use the same one1m Lglass syringe
note2:The syringe is preheated to the same temperature as the sample
note3:Pay attention to changes in the retention time of ethylene oxide each time to prevent gas leakage from the injection vaporization pad
note4:each sample(Including standard samples)Analyze three times in the shortest possible time, and two out of the three analyses must have results that do not differ by more than5%Otherwise, this sample should be reanalyzed.
5.2Draw the standard curve using the data measured by the standard sample(X:EOconcentration/(g/L); Y:Peak height or area).
5 .3Find the corresponding concentration of the sample from the standard curve. If the measured sample results are not within the range of the standard curve, the concentration of the standard solution should be changed to create a new standard curve.
6Calculation of residual ethylene oxide test results
The residual amount of ethylene oxide is expressed in terms of content or relative content.
According to the formula(H.1 )Calculate the content of ethylene oxide in the sample:
WEO= 5c1·G
in the formula:
WEO—Ethylene oxide content per unit product, measured in milligrams(mg);
5c1—The corresponding concentration of the test solution found on the standard curve, in grams per liter(g/L);
G—The quality of a unit product, measured in grams(9).
2According to the formula(H.2)Calculate the relative content of ethylene oxide in the sample:
CEO= 5c1*1 0 00
in the formula:
CEO—Relative content of ethylene oxide in the product, measured in milligrams per kilogram(mg/kg);
c1—The corresponding concentration of the test solution found on the standard curve, in grams per liter(g/L).
Our company also produces gas chromatographs; Liquid chromatograph; Chromatography workstation; Chromatographic column; Chromatographic column temperature chamber; Solvent filtration bottle vacuum pump; Ultrasonic cleaner; Chromatographic accessories; Sample bottle; Baijiu standard sample; Standard products; Turbidity meter; Acidity meter; Conductivity meter; Drying box; incubator; Electric furnace; Water quality analyzer; Food safety detector; Chromatographic accessories; Sample bottle; Baijiu standard sample; Standard products; Chromatographic reagents; Visible/UV spectrophotometer; Atomic absorption spectrophotometer; Infrared spectrophotometer, etc; Tungsten lamp; Element lamp; Colorimetric dish; Printing paper; Four color printing pen; Inkjet head; Chromatographic column; Deoxygenation tube; Injection needle; Switching valve; Steady flow valve; Pressure stabilizing valve; Electrode; Thin layer chromatography; Ion chromatography; YMC chromatographic column; TSK chromatography column, second-hand gas chromatograph, etc.
Wenzhou Ruixin Instrument Co., Ltd
Wenzhou Ruixin Instrument Co., Ltd. is engaged inA company that integrates sales, services, and maintenance of industrial testing instruments, physical and chemical analysis instruments, and measuring instrumentsWelcome to inquire and place an order,:
company:0 5 7 7-6 6 8 8 6 0 1 7
0 5 7 7-6 6 8 8 6 0 1 9
:rxin17
:http://www.rxin17.com/
:
Chen Jinwan 1 3 7 5 8 7 6 8 8 1 3:
Yang Yelong :