Purpose and CharacteristicsUse and characteristics
|
1 |
purpose |
◆Suitable for electrical, electronic, petrochemical Biotechnology, with various levels of refrigeration power output, can meet high-precision cyclic cooling and immersion tests for large industrial equipment to small experimental analysis instruments, including: laser equipment, electron microscopes, X-ray machines and X-ray machines, injection molding machines and press machines, welding machines and cutting machines, printing machines, material testing machines, nuclear magnetic resonance machines, linear accelerators, atomic absorption spectrometers, high-pressure disinfection axes, vacuum systems, jacketed experimental devices, fermentation devices, night pressure pumps, semiconductor industry equipment, rotary evaporators, battery equipment, viscometers, polarimeters, etc |
|
2 |
high-performance |
◆Adopting a fully enclosed high-pressure circulation pump for mixing, the temperature field is uniform and there is no leakage By integrating the sensor with the circulating pump flow, the stability control accuracy and temperature uniformity are greatly improved ◆This product is designed to be advanced and reasonable, and can adapt to long-term, stable, safe, and reliable testing requirements Easy to use and operate, with LCD soft buttons for a more concise and intuitive design |
|
Control system: spontaneously developed, LCD display Each control port can be directly connected to 220V voltage, electrical components, using high-quality domestic products or Schneider from France, Fuji from Japan, etc Ensure the security of the control system Additional functions: Fault alarm and cause prompt function, fault name can be changed, temperature upper limit protection function |
||
|
◆ Temperature controller: accuracy of 0.001 ℃ (display range) ◆ Temperature measuring body: PT100 testing sensor Control mode: Thermal balance temperature regulation mode, PID control |
IISpecificationIts main technical indices
|
order |
project |
content |
|
|
1 |
Studio Size |
300 x 350 x 300mm (depth x width x height); |
|
|
Box dimensions |
Approximately 780 x 510 x 900mm including caster size (depth x width x height); |
||
|
2 |
temperature range |
10℃~+70℃; (Extreme temperature -10 ℃~+80 ℃) |
|
|
temperature fluctuation |
≤±0.05℃; |
||
|
Horizontal temperature difference |
±0.05℃; |
||
|
heating up time |
10 →+70 ℃,<30min no-load average throughout the entire process |
|
|
|
cooling rate |
+70 → 10 ℃, approximately 1.0/min, average load throughout the entire process |
||
|
3 |
total power |
5KW |
|
|
4 |
Refrigeration capacity |
20 ℃ ≥ 2000W no-load |
|
|
5 |
internet traffic |
32L/min |
|