SW-TFG-12 type fume hood


The SW-TFG series fume hood is a commonly used equipment in laboratories that requires the discharge of harmful gases and the cleaning of river sewage during the experimental process.


SW-TFG-12Laboratory specific fume hoodworking principle

Integrated system of water, electricity, gas, and ventilation, equipped with multifunctional power sockets for easy testing of other electrical equipment during the experimental process. Adopting a quick opening valve facilitates the use of water during the experimental process. The front baffle is a movable glass door, and the top is a low-speed exhaust fan, which can smoothly discharge harmful and odorous gases during the experiment. The bottom of the working face is equipped with a stainless steel sink, which can discharge disinfectant and experimental residues from a row of sinks through water washing, maintaining a safe and reliable experimental environment.


SW-TFG-12Laboratory specific fume hoodStructural characteristics

The frame is made of 50% fireproof density board material, and the experimental area is covered with stainless steel veneer, which has good sealing, beautiful appearance, and strong durability.


Schematic diagram of SW-TFG-12 structure

technical parameter

model

SW-TFG-12

Average wind speed

0.25-0.45m/s

Inlet pressure

>0.5Pax

Power Supply

AC, Single phase 220V/50HZ (200W)

power waste

400W

weight

150KG

Working size

1030*695*580mm

External dimensions

1185*760*1950mm

working principle

The air is sucked into the static pressure box by the low-noise centrifugal fan at the top, and evenly blown out through the PVC pipe at the top (with the exhaust pipe connected to the outside), thus forming a negative pressure air curtain in the original experimental area, removing the original natural air or polluted air in the work area. After five minutes of operation, the ideal space can be reached and maintained. The air volume is adjusted by a touch switch, and the wind speed in the packaging work area is always in the ideal state.

Installation and use

The equipment should be installed in a relatively clean working environment. After plugging in 220V AC power, it can be used by turning on the functions shown on the control panel. Before starting up, the working area and shell of the fume hood should be carefully cleaned to remove surface dust. Normal operation and use can be carried out within five minutes after starting up.

maintain

1. In general, when using the working voltage adjustment point of the fan, if the ideal wind speed cannot be achieved, it indicates that there is a lot of dust accumulation in the air duct and it needs to be cleaned or updated carefully.

2. When replacing air ducts or accessories, attention should be paid to the correct model, specification, and size (original configuration), and ensure that the exhaust duct is sealed around the indoor area without any leakage.

General faults, causes, and troubleshooting methods

Fault phenomenon

reason

Discharge method

The main power switch cannot be closed and automatically trips

1. Fan jamming causes motor stalling or short circuit in the circuit

1. Adjust the position of the fan shaft, or replace the impeller and bearings, and check if the circuit is intact.

2. Check the insulation resistance of the circuit and components to the casing point by point according to the circuit diagram, and repair the insulation fault.

The fan is not turning

1. The contactor is not working.

2. The fuse of the fan power supply has blown.

1. Check if the contactor circuit is normal.

2. Replace the melt core.

The fluorescent light is not on

1. The lamp or relay is damaged.

2. The fuse of the lamp power supply has blown.

1. Replace the light tube or relay.

2. Replace the melt core.