One LC8245 four element low-pressure infusion pump
One LC8320 automatic sampler
1 LC8400 column temperature box
-
LC8500 UV Visible Detector 1 unit
The complete set of instruments, including workstations, complies with GMP/GLP and FDA CFR 21 PART 11 regulatory requirements!
(The specific indicators and parameters of this product are subject to the quotation)
1、 Infusion pump unit: Four element low-pressure infusion pump
★ 1. Maximum pressure resistance: 6000 psi (40Mpa)
2. Flow rate range: 0.001 to 10.000 mL/min (increments of 0.001 mL/min)
★ 3. Traffic accuracy: ≤ 0.06% RSD(ASTM)
4. Flow accuracy: ± 0.2%
5. Pressure pulsation: ≤ 1%
6. Gradient accuracy: ± 0.5%
7. Gradient repeatability: ≤ 0.1%
★ 8. plunger online cleaning: standard, high/medium/low speed
9. Channel volume: Four channel degassing function, 480ul/channel
2、 Detector: UV Visible Detector
1. Wavelength range: 190-800 nm
2. Light source: deuterium lamp, tungsten lamp
3. Spectral bandwidth: 8 nm
4. Wavelength accuracy: ±1 nm
5. Noise: ±0.35×10-5 AU(ASTM)
6. Drift: 1×10-4 AU/Hr
7. Wavelength accuracy: 0.2 nm
8. Linear range:> 2.5AU (ASTM)
9. Maximum sampling rate: 100Hz
10. Flow pool pressure resistance: 1200PSI
11. Flow pool optical path length: 10mm
12. Minimum detection concentration: 4×10-9 g/mL
3、 Injector: Automatic Injector
1. Injection mode: full loop injection/partial injection/microliter carrier injection
2. Quantitative ring volume: standard 100 μ L
3. Sample repeatability: Full cycle injection: RSD ≤ 0.3% Partial injection: RSD ≤ 0.5% microliter Carrying injection: RSD≤1%
4. Sample capacity: 2 standard 48 position sample bottle racks with 1.5mL injection bottles (standard)
(Optional 96 well plate, 384 well plate)
★ 5. Cross contamination: < 0.005%
6. Linear: 0.9999
7. Maximum pressure resistance: 6000psi (customizable)
4、 Column temperature box
1. Temperature control range: room temperature+5 ℃ -70 ℃
2. Temperature accuracy: ± 1.0 ℃
3. Temperature stability: ± 0.1 ℃
4. Column capacity: 1 250mm chromatographic column (can be equipped with an additional 10mm protective column)
5、 Chromatographic column
Standard C8250 × 4.6mm, 5um
6、 Chromatography workstation
1. Regulatory Applicability: Compliant with FDA 21 CFR Part 11/GMP/GLP specifications;
2. Usability: A graphical monitoring interface that allows for real-time monitoring of device status, making it simple and easy to use;
3. Data processing: powerful data processing capabilities, providing multiple integration functions and calculation methods;
4. Rich features: automatic peak width, automatic threshold algorithm, adaptive data collection, automatic adjustment algorithm to achieve the best integration effect;
5. User management: With powerful and comprehensive user management functions, access permissions can be set for easy management, and different usage permissions can be set according to actual needs;