KF-6901D is a portable 5-channel partial discharge inspection and testing instrument that can provide five detection methods according to different configurations: transient earth voltage (TEV), air ultrasonic (AA), contact ultrasonic (AE), high-frequency current (HFCT), and ultra-high frequency (UHF); The detection results display discharge intensity, discharge phase, historical data, etc. in numerical and graphical form, while providing environmental measurements of temperature, humidity, and air pressure values.
Functional Features
U has rich detection functions, providing multiple detection methods such as transient earth voltage (TEV), ultrasonic (AA and AE), high-frequency current (HFCT), ultra-high frequency (UHF), etc.
U provides test data management and allows for one click saving of test data.
The human-machine interaction adopts the latest ARM processor and embedded operating system, with touch operation and user-friendly interface.
U provides a USB interface for data export.
The built-in function self-test ensures that the instrument functions properly.
U can measure the environmental temperature and humidity of the detected object.
The attachment is rich and convenient for testing various devices, including TEV probes, ultrasonic parabolic waveform concentrators, contact ultrasonic probes, external extended microphone probes, HFCT sensors, ultra-high frequency sensors, non-contact infrared temperature sensors, etc.
According to ergonomic principles, it is convenient for handheld operation.
Application scenarios
It is suitable for live detection and positioning of partial discharge in electrical equipment such as distribution cabinets, switch cabinets, transformers, cables (including cable trays, joints, etc.) and insulators, pole mounted switches, GIS, etc. of overhead lines in substations and factories.
Hardware parameters
1.1.1 TEV transient ground voltage measurement
U Sensor: Capacitive
Measurement range: 0-60dBmV
Frequency range: 1-100MHz
U resolution: 1dB
Accuracy: ± 1dB
Maximum number of pulses/cycles: 1400
Minimum pulse rate: 10Hz
1.1.2 AA air ultrasonic measurement
Measurement range:- 7dBuV~68dBuV
U resolution: 1dB
Accuracy: ± 1dB
U Sensor sensitivity:- 65dB (0dB=1 volt/ubar effective value SPL)
Sensor center frequency: 40kHz
Frequency range: 20-100kHz
Sensor diameter: 16mm
U heterodyne frequency: 38.4kHz
1.1.3 AE Contact Ultrasonic Measurement
Measurement range: 0dBuV~68dBuV (0.001mv~1.000mv)
U resolution: 1dB
Accuracy: ± 1dB
U Sensor sensitivity:- 65 dB (0 dB=1 volt/μ bar effective value SPL)
Frequency range: 0-200 kHz
1.1.4 HFCT high-frequency DC measurement
U Sensor: HFCT
Measurement range: 0pC to 25000pC
Sensitivity: not less than 50pC
Detection frequency: 3MHz~30MHz
Inner diameter: 45mm
Outer diameter: ≥ 110 mm
Output impedance: 50 Ω
Output interface: BNC connector
1.1.5 UHF ultra-high frequency measurement
U Sensor: UHF
Measurement range:- 75dBm ~0dBm
Sensitivity:> 12mm equivalent height
Resolution: ± 0.5dB
Frequency response: 300MHz~1500MHz
Output impedance: 50 Ω
Output interface: BNC type
1.1.6 Input and output interfaces
U LEMO external ultrasonic sensor input interface
BNC external AE sensor input interface
BNC external HFCT sensor input interface
BNC external UHF sensor input interface
3.5mm stereo headphone jack
U standard SD card data storage interface
2.1mm Low Voltage DC (18VDC) Charger Interface
1.3mm power supply synchronous input interface
USB interface
RJ45 Ethernet interface
1.1.7 SD card
U comes standard with 16GB (supports 4GB and above SD card)
1.1.8 Earphones
3.5mm stereo headphone jack, 32 ohms
1.1.9 Built in power supply
Internal battery: 3.7V/31.2Ah lithium-ion battery
Normal working hours: approximately 8 hours
Battery protection function: automatic alarm when low voltage occurs