ISDC Body Temperature Calibration Blackbody - Thermal Measurement Instruments - Radiation Temperature Measurement - Nightingale Blackbody Product Description:
ISDC Body Temperature Calibration Blackbody Thermal Measurement Instrument Radiation Temperature Measurement Nightingale Blackbody is a new low-cost solution provided by ISDC and Santa Barbara Infrared for real-time body temperature measurement systems. The Nightingale blackbody system is specifically designed for on-site comparison of human body temperature, and the ISDC body temperature calibration blackbody provides a highly stable, uniform, and easy to operate thermal radiation source. Using ISDC temperature calibration blackbody in the same scene as the subject can immediately visually and report the temperature of the infrared camera. Infrared cameras and position sensors are affected by many factors in the environment, and ISDC temperature calibration blackbody can minimize their impact. The ISDC body temperature calibration blackbody has a compact size of only 1.5 pounds, making it very easy to deploy. A black body is an idealized object that can absorb all external electromagnetic radiation without any reflection or transmission. In other words, a blackbody has an absorption coefficient of 1 and a transmission coefficient of 0 for electromagnetic waves of any wavelength. Physicists use this as a standard object for studying thermal radiation. It can completely absorb all external electromagnetic radiation without any reflection or transmission, and this object is called an absolute blackbody, abbreviated as a blackbody.
Infrared Systems Development Corporation (ISDC) was founded in 1997 and acquired Graseby Infrared's instrument product line from Infrared Industries (IRI), Infrared Systems (IRS), and Barnes Engineering Company. These products include blackbody sources, spectral radiometer systems, energy modulators, and accessories. Subsequently, ISDC improved and enhanced many of its products, and developed various new products such as infrared detector preamplifiers and TE cooled blackbodies. All products are developed and manufactured at our factory located in Orlando, Florida. ISDC also develops customized application systems and provides design consulting services for many infrared instrument and subsystem projects

A black body is an idealized object that can absorb all external electromagnetic radiation without any reflection or transmission. In other words, a blackbody has an absorption coefficient of 1 and a transmission coefficient of 0 for electromagnetic waves of any wavelength. Physicists use this as a standard object for studying thermal radiation. It can completely absorb all external electromagnetic radiation without any reflection or transmission, and this object is called an absolute blackbody, abbreviated as a blackbody
Application scope:
- Temperature measurement system detection and calibration
- Thermal imager calibration
Product features:
- Real time monitoring
- Compensate for environmental impacts
- Eliminating thermal imager errors
- Calibrate thermal imager
- Equipped with humidity and environmental sensors
ISDC temperature calibration blackbody - thermal measurement instrument - radiation temperature measurement - Nightingale blackbody technical parameters:
Temperature range: 30–45oC (86-113oF)
Wavelength range: 3–14 um
Emissivity:> zero point nine five
Transmitter size: 3” x 3” (76.2 x 76.2 mm)
Radiation source type: surface type
Temperature uniformity:+/- zero point one fiveoC (+/- 0.3oF), Center 1.5 x 1.5 "area
Calibration accuracy:+/- zero point one fiveoC (+/- 0.3oF) Radiation intensity
Temperature stability:+/- zero point zero fiveoC (+/- 0.1oF)
Response time:< 5 Min
Set temperature resolution: zero point oneoC (0.2oF)
Control type: active multi band P.I.D
AC adapter line voltage: 90-125 or 208-240 VAC 50-60 Hz
Power requirement: Provide 18 VDC, 1A AC power adapter
Operating environment: 22° C +3/- 5°C (71.6°F +5.4/- 9°F)
5%-90% R.H. N.C.
Storage temperature range:- 20°C to 70°C (-4°F to 158°F)
Warranty period: 1 year
Remote interface: USB
Base: ¼ -20 tripod
Weight: 1.5 lbs (0.682 Kg)
model: BTR-03