Beijing Huaxu CenturyThe HXMF03 series gas mass flowmeter/controller adopts capillary thermal sensing technology and the principle of thermal temperature difference, that is, the heat exchange relationship between the fluid and the heat source in the flow, to measure and control the mass flow rate of gas.
Beijing Huaxu CenturyThe HXMF03 series gas mass flowmeter/controller adopts capillary thermal sensing technology and the principle of thermal temperature difference, that is, the heat exchange relationship between the fluid and the heat source in the flow, to measure and control the mass flow rate of gas. Since the thermal temperature difference is only related to the mass flow rate of the gas, its measurement value is not affected by changes in gas temperature and pressure.
The HXMF03 series of medium flow thermal mass flow products adopts a compact design, with a small and exquisite volume, and the product is sturdy and durable. The product uses a linearization correction circuit and EMI protection, which has the characteristics of safety, reliability, stable performance, low power consumption, and long service life.
In terms of data transmission, it not only has the traditional wired type, but also has powerful functions such as wireless data transmission through LAN, Internet remote data monitoring cloud platform, etc. Can effectivelySatisfy the different working conditions of various customers, widely used in many industrial and scientific research fields.
Product Features:
● quick response
● Ultra high cost-effectiveness
Accurate measurement and excellent long-term stability
Measurable corrosive gases, sturdy and durable
Exquisite appearance design, comprehensive safety protection circuit and EMI
● Can be equipped with a dedicated display meter to show instantaneous flow rate
● Multiple options: high precision, custom range, wireless transmission of flow signals
Remote data monitoring cloud platform, SMS alarm, email alarm, etc
Typical applications:
· 光伏及晶体制备·Vacuum industry and coating/leak detection·Furnace and microreactor device
·Semiconductor industry, universities/research institutes·Analytical instruments · Gas mixing ratio
·Measurement and Pricing·Fuel cells, combustion testing, and other gas measurement and control applications
Performance indicators:
Media requirements:Gas compatible with the medium, supporting corrosive gases
Type of medium:Customer specified, can also be changed on-site-Need to consult us
Calibration condition:0 ℃ @ 1 atm (nitrogen)
Measurement and control range:fromChoose from 0-30slm to 0-250slm, and customize a higher flow range
Range ratio:50:1
Accuracy:± 2% Full Scale High Precision Options: ± 1% Full Scale, ± 1% Reading
Repeatability:± 0.2% full scale
Traffic resolution:0.1% full scale
Response time:2S
Valve status:Normally closed (controller only, flow meter without valve)
Operating Temperature5~45℃
Working humidity:0-100% RH (non condensing)
Working pressure difference:flow meter0.02~2MPa(≤100slm),0.05~2MPa(>100slm)
controller0.1~0.3MPa(≤100slm),0.15~0.3MPa(>100slm)
Maximum withstand voltage:2MPa
Input and output signals:The standard is4~20mA
optional1~5V、0~5V、Modbus、RS485
Supply voltage:flow meter24VDC@100mA , 控制器 24VDC@400mA
Electrical interface:DB 15pin
Leakage rate:≤1×10-9powerPa·cubic meter/S(He)
Liquid receiving material:316L SS
Sealing material:According to the type of gas selected by the manufacturer or specified by the customer, including:EPDM、fluororubberFKM、 丁腈橡胶 NBR、 氯丁橡胶CR、硅橡胶VMQ
Pipeline joint:3/8 ", 1/2" double card sleeve, or 8, 10, 12mm double card sleeve
Overall dimensions:Length, width, and height: flow meter157 × 40 × 125mm, controller 182 × 40 × 125mm
Product weight:flow meter1.16kg controller2.26kg
Order Information:
· Flow meter or controller·Gas type and flow range·Working temperature
·Inlet pressure and outlet pressure·Pipeline connection size and type
·Input/output signal·Installation posture, options, and other special requirements: Please consult us