Product Introduction:
Rich features - meeting customers' different analysis needs
The hardware has VP function, which records maintenance information and operation records, in compliance with GLP/GMP requirements; The system controller is equipped with user-friendly design features such as a clock, thermometer, and hygrometer.
Excellent performance - meeting customers' strict requirements for instruments
The detector adopts a dual beam monochromator composed of imported deuterium lamps, photovoltaic cells, and 1200/mm concave gratings; The precision machined dual lens flow cell, high-precision microprocessor for wavelength adjustment, and dual channel high-speed sampling frequency ensure low noise, low drift, and ultra-high sensitivity.
The VI2010 workstation complies with various regulatory requirements.
Reliable results - meeting customers' accurate requirements for results
Comparative experiments were conducted with imported instruments, and the analysis results showed a high degree of consistency.
Easy operation - convenient for customers to proficiently operate the software
The software adopts a multi window mode, which is easy to operate.
Exquisite appearance - meeting customers' visual requirements for instruments
Exquisite appearance brings visual enjoyment.
Technical indicators and parameters:
|
LC-10TvpHigh pressure constant current infusion pump
|
|
|
Infusion method |
Micro volume series double plunger |
|
Maximum infusion pressure |
0~6000Psi |
|
Flow setting range |
0.001-9.999ml/min (flow rate adjusted in 0.001ml/min steps) |
|
Error in flow setting value |
≤0.5% |
|
Flow stability error |
≤0.2%RSD |
|
pressure pulsation |
Less than 15Psi (flow rate 1mL/min, pressure 600-1600Psi). ) |
|
Pump sealing performance |
The pressure is 5400Psi, the time is 10 minutes, and the pressure drop is less than 400Psi. |
|
Time program function |
have |
|
size |
W260×H130×D420mm |
|
weight |
11kg |
|
Range of environmental temperature for use |
4~40℃ |
|
SPD-10TvpUV Visible Variable Wavelength Detector
|
|
|
wavelength range |
190nm~700nm |
|
Wavelength indication error |
≤±1nm |
|
Wavelength repeatability error |
≤±0.1nm |
|
Dynamic noise |
≤ ±0.75×10-5AU (甲醇,1ml/min,254nm,20℃ 。) |
|
Static noise |
≤ ±0.5×10-5AU (Empty Pool, Response Time 1 second, 20 ℃) |
|
Dynamic baseline drift |
≤±1×10-4AU/h (甲醇,1ml/min,254nm,20℃ 。) |
|
Static baseline drift |
≤0.5×10-4 (Empty pool, response time 1 second, 20 ℃.) |
|
linear range |
≥104 |
|
Minimum detectable concentration |
≤1×10-9G/mL (naphthalene/methanol solution) |
|
Qualitative repeatability |
RSD6≤ 0.1% |
|
Quantitative repeatability |
RSD6≤ 0.5% |
|
spectral bandwidth |
6nm |
|
Flow pool volume |
8μL |
|
optical path |
10mm |
|
Time program function |
have |
|
size |
W260×H130×D420mm |
|
weight |
11kg |
|
Range of environmental temperature for use |
4~40℃ |
Standard configuration:
| LC-10Tvp high-pressure constant current pump: | 2 units |
| SPD-10Tvp UV detector: | 1 unit |
| Tray: | 1 piece |
| Mixer: | 1 piece |
| 8725i manual injection valve: | 1 can be optionally equipped with 7725i manual injection valve |
| Chromatography workstation: | 1 set |
| Liquid chromatography column: | 1 piece (C18 4.6 * 250mn, 5um) |
| Micro injector: | 1支 (50ul/100ul) |
| Reverse pressure valve: | 1 piece |
| Injection bracket: | 1 (for injection valve) |