model
|
VLPT360
|
VLPT720
|
VLPT1000
|
VLPT2370
|
|
Volume liter (L)
|
360
|
720
|
1000
|
2370
|
|
Studio ruler
Inch(mm)
|
deep
|
800
|
1000
|
1000
|
1400
|
|
wide
|
600
|
800
|
1000
|
1300
|
|
tall
|
750
|
900
|
1000
|
1300
|
|
Overall dimensions
(mm)
|
deep
|
1900
|
2200
|
2200
|
1650
|
|
wide
|
1100
|
1310
|
1510
|
1810
|
|
tall
|
1900
|
2050
|
2150
|
2450
|
|
temperature range
|
LPT: -70~+160℃ LPTH:(ambient temperature+20)℃~+160℃
|
|
humidity range
|
10~98%R.H
|
|
pressure range
|
atmospheric pressure~0.5kpa
|
|
temperature fluctuation
|
≤± 0.5 ℃ (at atmospheric pressure)
|
|
Temperature Uniformity
|
≤± 2 ℃ (at atmospheric pressure)
|
|
Humidity fluctuation
|
≤± 2% R.H (at atmospheric pressure)
|
|
Humidity uniformity
|
≤± 3% R.H (when relative humidity>75% R.H); ≤± 5% R.H (when relative humidity ≤ 75% R.H)
|
|
pressure deviation
|
≤±2Kpa≥40 kpa),≤±5%(2~40kpa)time),≤±0.1Kpa≤2Kpatime)
|
|
warm-up time
|
25℃(At room temperature)→+160℃ ≤ 60 minutes (normal pressure)
|
|
cooling time
(Limited to)LPTtype)
|
+25→-65℃(At normal pressure)
|
|
≤ 40min (normal pressure)
|
≤ 60 minutes (normal pressure)
|
≤ 90 minutes (normal pressure)
|
|
Voltage reduction time
|
atmospheric pressure~1.0 kpa(At room temperature and when dry inside the box)
|
|
≤20min
|
≤25min
|
≤30min
|
≤45min
|
|
Temperature and humidity regulation method
|
BT(H)CBalanced temperature and humidity control method
|
|
Environmental temperature for use
|
+5℃~35℃
|
|
material
|
shell
|
Double sided galvanized steel sheet,Surface spraying treatment,Shell pressure bearing
|
|
inner body
|
stainless steel plateSUS304
|
|
adiabatic
|
Vacuum specific composite pressure insulation material
|
|
tune
section
device
|
heater
|
Air heater; Nickel chromium alloy electric heater; Wall heater; Stainless steel armored electric heater
|
|
humidifier
|
Stainless steel armored electric heater
|
|
fan
|
Magnetic coupling driven axial fan
|
|
refrigeration compressor
|
Low noise piston compressor
|
|
Cooling method
|
Water cooled mechanical compression binary cascade refrigeration
|
|
refrigerant
|
R404A/R23
|
|
controller
|
operating interface
|
8inch,640x480dot matrix,TETTrue color touch screen
|
|
Operation mode
|
Fixed value method, program method
|
|
Program memory capacity
|
editable:20A program,Maximum for each program50step,Can be set20A cycle(Maximum per cycle step9999Next cycle)fixed, tenA program can be linked,Can achieve infinite operation.
|
|
Set indicator range
|
temperature:Adjust according to the temperature range of the equipment(upper limit+5℃,lower limit-5℃)humidity:0~100%R.H(Equipment humidity) Air pressure: 0~100pcs
|
|
display resolution
|
temperature:0.1℃;time:0.1minhumidity:0.1%R.H(Equipment for humidity control Pressure: 0.01kpa
|
|
input
|
TType thermocouple,4~20mAcurrent
|
|
communication function
|
(select)RS-485interface,Capable of local and remote communication(Centralized monitoring software is requiredRS-485/RS-232Conversion, requires occupationPCMachine'sCOMMouth andUSBOne mouth each)Can be connected simultaneously at most16Equipment, maximum cumulative cable length800m,PCSoftware usage environment:IBM PCDual use machine,PIIaboveCPU,128MThe above memory,Simplified ChineseWindows2000Or Simplified ChineseWindowsXPoperating system.
|
|
Control method
|
Anti integral saturationPID
|
|
Curve recording
|
Equipped with battery protectionSRAM,Can save the set values, sampling values, and sampling time of the device, and can be observed by sliding in real-time with finger touch on the screen;The default maximum storage time is60sky(The sampling period is1.5min),The formula for calculating sampling time is as follows:If the sampling period is set tot/min,storage timed=((65536xt)/60)/24(sky).
|
|
data collection
|
*Screen interface curve output--Directly save the curved interface on the touch screen or any interface requested by the customer as*.bmpSave image toUPrint the curved interface on the touch screen or customer requirements directly from the disk or printer(Need to modify the interface software)Any interface;
*USBfunction--match1GUSB flash drive1a.During the experimentUInsert the disk into the controllerUSBOn the interface,The experimental data will be automatically recordedUInside the plate,Save as*.csvdata file(Non curve chart),The sampling period can be set on the screen.useEXCELWait for the software to open and process;
*Connect to printer(select):-configure printer,The test date can be printed by setting the number of printing times according to time,set temperature,Actual measured temperature,Set humidity,Measured humidity;
*pick upPCmachine(select):--Through centralized monitoring software,Can record experimental data,inPCAutomatically display as a curve in the machine,Can be printed directly,Unrestricted recording time.The file size depends on the hard drive capacity.PCIt can also be used as an operating terminal,Realize remote monitoring.
|
|
Supplementary functions
|
Fault alarm and cause, processing prompt function;A fail-safe mechanism;Upper and lower limit protection function;timer function(Automatic start and automatic stop operation).
|
|
供水
|
Water supply method: water pump lifting Water storage tank: 50L Water quality: electrical resistivity>500 Ω m
|
|
power supply
|
AC380(1±10%)V (50±0.5)Hzthree-phase four-wire+Protect the ground wire
|
|
Installed capacity (KW)
|
LPT:15 LPTH:8
|
LPT:20 LPTH:10
|
LPT:20 LPTH:10
|
LPT:28 LPTH:18
|
|
Standard Configuration
|
Multi layer electric insulated glass observation window1a,power cord(long5rice)1article,lighting lamp1a,caster4a,Sample power control terminal1a,Uplate1a,One external pressure testing interface,Multi action inflatable brakeDN25One, sample shelf2layer
|
|
safety device
|
Refrigeration system: compressor overpressure protection, compressor motor overheating, compressor motor overcurrent, cooling water supply undervoltage, adjustable insulation protection, test space temperature fuse, air conditioning channel extreme overtemperature protection, heater overcurrent fast fuse, wall heater extreme overtemperature protection. Test box and fan motor overheat protection.
Other: Vacuum pump motor overcurrent, overload protection, main power phase sequence and phase loss protection, leakage protection
|
|
meet the standard
|
GB/T2423.25, GB/T2423.26, GJB150.2, GB/T10591, GB/T10592
|