The HCS-5000 infrared carbon sulfur analyzer is an intelligent infrared analysis instrument composed of a high-frequency induction combustion furnace and a microcomputer control system, with a pyroelectric sensor as the core. The analysis software is based on the WINDOWS XP operating platform, with a standard WINDOWS Chinese operating interface and user-friendly human-computer interaction functions. Mainly used for the analysis of carbon and sulfur content in ferrous metals, non-ferrous metals, rare earth metals, inorganic substances, ores, ceramics and other materials in industries such as metallurgy, machinery, commodity inspection, scientific research, and chemical engineering.
Main Technical Performance
Adopting humanized design, the structure and control board are all produced using modern processing technology, and the products are meticulous, precise, beautiful and generous
The microprocessor unit adopts the most popular ARM9 system and integrates dual communication interfaces of high-speed USB and Ethernet TCP/IP protocols
The first domestically developed linearization calibration software that utilizes multivariate nonlinear fitting techniques, including single point fitting and multi-point fitting correction
Adopting high-performance lithium tantalate pyroelectric infrared sensors to improve system detection sensitivity
Constant pressure and current in the pneumatic system, analysis of data stability
Using 3 5KW air-cooled ceramic power tubes, the heating power is stable and reliable
High frequency furnace power adjustable, suitable for different material sample analysis requirements
The automatic cleaning device of the furnace head can reduce the impact of dust on the analysis results
The furnace head heating device makes the conversion rate of sulfur tend to be consistent, improving the stability of sulfur determination
Comprehensive database retrieval, remote network query of analysis results
Dual carbon and dual sulfur gas chambers are optional
Main technical parameters of HCS-500 infrared carbon sulfur analyzer
1. Analysis scope:
Carbon: 0.00001% to 15% (expandable to 99.99%)
Sulfur: 0.00001% to 5% (expandable to 99.99%)
2. Sensitivity (minimum reading): 0.000001%
3. Analysis accuracy (repeatability):
Carbon: RSD≤0.5%
Sulfur: RSD≤1.0%
4. Analysis error:
Carbon meets ISO9556-94 standard
Sulfur complies with ISO4935-94 standard
5. Analysis time: 20-40s (automatic control)
6. Electronic balance: 0-10g
Weighing accuracy: 0.0001g