Hash 1950 Plus TOC AnalyzerTypical applications:
Used for monitoring TOC content and determining TOC removal rate in surface water and tap water. Compliant with standard method 5310C and EPA method 415.1.
Instrument features:
● Automatic dilution function, automatic calibration, automatic cleaning
Built in intelligent software that can automatically display TOC removal rate
● Dual channel sampling analysis
● Manual sample capture analysis
Nitrogen or air can be used as the carrier gas (through a gas purifier)
● Automatic diagnosis, no water sample, no reagent, no gas alarm
● Complies with USEPA415.1, standard method 5310C, and ISO8027 standard analytical methods
Analysis principle:
The instrument adopts ultraviolet light/persulfate oxidation method.
The sample is first acidified, oxidizing the total inorganic carbon (TIC) in the water to carbon dioxide; Then use a carrier gas to blow out the carbon dioxide; Then add persulfate to the water sample and oxidize it with ultraviolet light (UV). At this time, TOC in the water is always oxidized to carbon dioxide; Then use a carrier gas to bring carbon dioxide into an infrared detector (NDIR) to detect the concentration of carbon dioxide and convert it into TOC concentration.
Hash 1950 Plus TOC AnalyzerTechnical indicators:
Measurement range: 0~5/10/25mg/L
Response time: T90 ≤ 8 minutes
Accuracy: ± 2% of full scale
Detection limit: When the temperature is 25 ℃ and the measurement range is 0-5mg/L
≤ 0.015mg/L
Injection conditions: 0.15-6bar (2-87psi);
Flow rate, 25-200mL/min
Operating temperature: 5-40 ℃
Sample temperature: 2-70 ℃
Data communication: RS232/RS485 serial port (optional)
Analog output: Two channels of 4-20mA (can be set as TOC concentration or TOC removal rate)
Power requirements: 115/220V AC, 50/60Hz
Instrument specification: 54kg, 981 × 675 × 220 mm
Ordering Guide:
1950 Plus online TOC analyzer, standard configuration includes: host, reagents, installation accessories, operation manual, etc
H-4195-1010-EPA 1950Plus, Range 0~5mg/L, dual channel
H-4195-1020-EPA 1950Plus, Range 0~10mg/L, dual channel
H-4195-1030-EPA 1950Plus, Range 0~25mg/L, dual channel
4P95-2100-00 standard chassis, cold-rolled steel shell, with visible window or optional other chassis. Please consult Hash's various offices for specific models
Optional attachments:
120161 TOC Installation Rack Component
4200-1001 PS200 sample filtration system, 100 μ m (please consult the department for other filtration specifications)
Spare parts package for the year 200123
Spare parts package for the year 200124 2
4300-0005 air purifier (without air compressor)
5920000 internal expansion card MODBUS (RS232)
5920001 internal expansion card MODBUS (RS485)