1Main technical indicators of the equipment:
1.1 measuring range(1gsample):0.00001%-99.999%
1.2 Analysis accuracy:RSD≤1%
1.3 Analysis error: compliantGB/T20123-2006/ISO15350:2000standard
1.4 Minimum reading:0.1ppm
1.5 Analysis time:20-60Adjustable in seconds, usually within35About seconds
1.6 Minimum reading:0.1ppm
1.7 Electronic balance reading accuracy:0.0001gAutomatic online input
1.8 Working environment: Indoor temperature:10-30℃Relative humidity: less than90%
1.9High frequency furnace: power≥2.7kVAFrequency:20MHz
1.10Power supply: requires good grounding
voltageAC220V±5%; Frequency50Hz±2%
1.11 Oxygen: Purity≥99.5%
INPUT0.18MPa±5%
Oxygen carrying pressure0.08MPa±2%
1.12 Power gas: nitrogen or compressed air (purified); INPUT0.25MPa±5%
1.13 Gas flow rate: Top oxygen flow rate1.0-2.0L/minAnalyze air flow rate3.0-4.5L/min
1.14 Desiccant: Efficient color changing desiccant
1.15 Filter agent: quartz cotton
2Main features of the equipment
-Circuit design: The whole machine adopts dualCPUModular design of upper and lower computers, selection of lower computersAtmega162To control the microcontroller. Highly integrated electronic circuits, stable and reliable; Simultaneously adopting multi-level concealed isolation circuits to completely solve high-frequency interference.
-Light source: Imported platinum infrared light source, with continuous heating and high spectral efficiency.
-Analysis pool: gold-plated carbon sulfur analysis pool and high-precision pyroelectric infrared detector (capable of10-11)Ensure signal stability.
-Motor: Digital modulation motor, good thermal stability, continuous service life10More than 10000 hours.
2.2 High frequency combustion system
-High frequency circuit: high-power high-frequency circuit design, using7.5KVAHigh frequency power tube (actual usage power greater than)2.7KVATo reduce the load on the high-frequency combustion system and improve its service life, no sample air burning is required.
-High frequency control circuit: automatically detects the working conditions of solenoid valves, lifting cylinders, and high frequency, automatically alarms for over time and over-current, and protects the high frequency furnace from working in a normal state.
-Air circuit: High precision electronic flow controller and imported air circuit system (solenoid valve, pipe joint, lifting cylinder) ensure stable air flow and automatic leak detection.
-Double calibration of fixed and gas standards.
-Dust removal system: The furnace head heating device ensures constant temperature of the furnace head; The buckle type automatic cleaning device can reduce the impact of dust on the analysis results; Imported high-pressure ash discharge system.
-0.4Micro sized ultra microporous metal filter ensures complete separation of dust and gas.
-Using current/voltage/power/Choose a method to adjust the furnace temperature, suitable for samples of different materials.
2.3 Random software
-WINDOWSFull Chinese operation interface, easy to operate and master;
-The program design adoptsDeliphSoftware compilation;
-The software provides a multi-user management system, which allows administrators to set different user permissions;
-Provide channel management function, allowing channels to be added, deleted, and edited without any quantity limit;
-Provide sample management function, which allows editing of sample names and labels, as well as adding and deleting sample names;
-The analysis results are based onACCESSDatabase storage, capable of storing all data and curves for each analysis of carbon and sulfur;
-The analysis results can be freely queried based on various factors such as time, operator, sample name, and identification;
-The switch diagnostic function can be implemented by software to detect the sealing of the furnace head and gas chamber;
-Sulfur curve box, dynamically displaying real-time data and sulfur release curves during the analysis process;
-Integration of dynamic data during the analysis process, per second24Subsampling improves the sensitivity and accuracy of analysis;
-Diversified printing modes, providing two printing modes for laboratory and testing stations, and the ability to design printing formats independently;
-The software is fully functional and provides curves/Multiple functions including data storage, blank deduction, parameter setting, channel selection, mathematical statistics, result correction, curve comparison, etc;
-Adopting high-speed technology24positionADS1224Sampling chip, with up to four sampling channels, can switch between sulfur and high sulfur detection pools at will;
-Pressure regulation and flow control can be achieved through software;
-Data can be remotely transmitted through networked operations, enabling furnace front reporting.