HS32-YQ3000-C Smoke and Gas Analyzer | ||||
Product Overview |
HS32-YQ3000-C Smoke and Gas AnalyzerThe application of pitot tube constant velocity sampling weight method to capture particulate matter in pipelines, and the use of constant potential electrolysis method for qualitative and quantitative determination of harmful gases, to achieve simultaneous detection of smoke and dust. It can be used by departments such as environmental protection, health, labor, safety supervision, military, scientific research, and education to determine the emission concentration/total amount of smoke and dust (gas) from various boilers and furnaces, as well as the dust removal and desulfurization efficiency of equipment. |
|||
|
1 YQ3000-C=1 traditional smoke and dust testing instrument+2 DC power supply boxes |
|||
Adopting standards |
HJ/T 48-1999 "Technical Conditions for Dust Samplers" JJG 968-2002 "Smoke Analyzer" JJG 680-2007 "Smoke and Dust Sampler" |
|||
|
Product advantages |
HS32-YQ3000-C Smoke and Gas AnalyzerProduct advantages: ◆ Built in replaceable battery, can be used without AC power supply; ◆ Optional tablet computer for remote operation; Adding a flue PM2.5 sampling gun can achieve PM2.5 detection in the flue; |
|||
|
Features: ◆ Multi purpose (capable of measuring smoke, exhaust gas, oil fume, and asphalt smoke) ◆ Built in replaceable battery, can be used without AC power supply High performance and long-life smoke sampling pump, with negative pressure up to 60kPa ◆ Efficient smoke and dust filtration device, with an accuracy of up to 1 μ m, to avoid contaminating the smoke and dust sampling pump ◆ Constant flow extraction of flue gas, with more stable and accurate measurement values ◆ Large capacity data storage (20000 sets), printing options available ◆ Sturdy casing, suitable for use in harsh environments ◆ Full process PTFE sampling to reduce smoke adsorption and improve measurement accuracy ◆ Automatic temperature/pressure correction, direct reading of standard condition data ◆ Dust collection and anti backflow function ◆ User password protection |
|||
|
technical indicators |
Main parameters |
Parameter Range |
Resolution |
Accuracy |
sampling flow |
(5~80)L/min |
0.1 L/min |
Better than ± 2.5% |
|
sampling flow |
Better than ± 2% (voltage varies between 180V and 250V, resistance varies within 3-6 kPa) |
|||
Smoke dynamic pressure |
(0~2000)Pa |
1 Pa |
Better than ± 2% |
|
Smoke static pressure |
(-30 to +30)kPa |
0.01kPa |
Better than ± 4% |
|
Front pressure of flowmeter |
(-50~0)kPa |
0.01kPa |
Better than ± 2% |
|
Temperature before flowmeter |
(-55~150)℃ |
0.1℃ |
Better than ± 2 ℃ |
|
Constant suction velocity |
(1~45)m/s |
0.1m/s |
Better than ± 2.5% |
|
flue gas temperature |
(0-500) ℃ (expandable) |
1℃ |
Better than 3.0 ℃ |
|
excess air coefficient |
0~99.99 |
0.01 |
Better than ± 2.5% |
|
O2 (optional) |
(0~25)% |
0.1% |
Indication error: not exceeding ± 5%; Repeatability: ≤ 2%; Response time: ≤ 90s; Stability: Within 1 hour, it refers to a change of ≤ 5% |
|
SO2 (optional) |
(0~5700/14000★)mg/m3 |
1 mg/m3 |
||
NO (optional) |
(0~1300/6700★)mg/m3 |
1 mg/m3 |
||
NO2 (optional) |
(0~200/2000★)mg/m3 |
1 mg/m3 |
||
CO (optional) |
(0~5000/25000★)mg/m3 |
1 mg/m3 |
||
CO2 (optional) |
(0-50)% (expandable) |
0.1% |
||
H2S (optional) |
(0~300/1500★)mg/m3 |
1 mg/m3 |
||
Constant speed tracking response time |
≤20s |
|||
Gas sensor lifespan |
About 2 years in the air |
|||
Overall dimensions |
(Length 420 x Width 220 x Height 340) mm |
|||
Total weight |
About 8.0kg |
|||
working power supply |
AC220V±10%, 50Hz |
|||