HIOKI 3194 Motor/Harmonic Testing Analyzer
The analysis equipment can be extended to a single instrument for motor evaluation, widely measuring power, speed, torque, conversion efficiency, and
Product details
Expanding analysis equipment to motor evaluation
Widely measure power, speed, torque, conversion efficiency, and harmonics with a single instrument
3194 can achieve evaluation of 3-phase variable frequency motor
By using the external signal input unit 9603-01, a torque sensor (strain gauge) can be directly connected to channel A. Then, the output of the tachometer (analog signal or pulse signal) can be input to channel B to obtain a system for measuring torque, revolutions per minute, and motor power.
Basic parameters
| Basic parameters | |
| Connection | Single phase two-wire to three-phase four wire (using different input units) |
| Measurement project | (When selecting 960096019602) Voltage, current, peak voltage/current, active/apparent/reactive power, phase angle, frequency, current/power accumulation, efficiency, load factor, power factor Harmonic (up to 3000 times), waveform (Use option 9603-01, input port should be used in conjunction with strain gauge torque sensor) Voltage, torque, RPM, Frequency, motor output |
| Measurement range |
Voltage:6.0000V~1.0000kV, (Depending on the input unit) electric current:200.00mA~500.00A, (Depending on the input unit) Power:Depending on the combination of input units for voltage and current |
| Integral range | 0 to ± 999999 TAh/TWh, (maximum integration time up to 10000 hours) |
| Basic accuracy Use 9600-9602 input units |
± 0.1% rdg. ± 0.1% f.s (voltage, current, power, between 45-66Hz) Note:When 9270-9272 and 9277-9279 are used simultaneously, the result accuracy is the sum of the accuracies of these clamp sensors mentioned above |
| frequency characteristic |
Using 9600:DC,0.5Hz~1MHz Using 9601:5Hz~100kHz Using 9602:DC 0.5Hz~200kHz (depending on the accuracy of the measuring clamp) |
| signal output |
Analog level:Voltage, current, active power, 5V DC f. s Waveform monitoring:Voltage, current, 1V rms f.s D/A output:Choose any output of 8 items, DC ± 5V f.s |
| Other functions | 6.4-inch TFT color LCD display screen, GP-IB/RS-232C interface scanning, average, FDD, True effective value/average rectification |
| power supply | 100/120/200/230 V AC (automatic switching), 50/60 Hz |
| Volume and weight | 430 wide x 150 high x 370 thick mm, 15kg (including all options) |
| attachment | Power line (1), connection line (1), voltage line (1) |
Options:
Input unit (3194 cannot be used alone, one or more input units need to be configured for measurement)
9600 AC-DC/DC direct input unit
9601 communication direct input unit
*9602 AC/DC clamp type input unit
9603-01 External Signal Input Unit
When selecting 9602 for current measurement, a clamp sensor is required
*9270 clamp current sensor 20 A AC
*9271 clamp current sensor 200 A AC
*9272 clamp current sensor 20/200 A AC
9277 universal clamp CT 20 A AC/DC
9278 universal clamp CT 200 A AC/DC
*9279 universal clamp CT 500 A AC/DC
*9290 clamp power converter 1000 A AC
printer
9604 printer unit
9232 recording paper (10m, 10 rolls/box)
*220H paper roll machine paper width 70-220mm
PC communicator
9151-02 GP-IB cable (2m)
9151-04 GP-IB cable (2m)
Online inquiry