HG112Low resistance meter testerIt is a standard for measuring, calibrating, and verifying DC low resistance testers (digital bridges, micro ohmmeters, etc.).HG112 low resistance meter testerThere are ten measuring ranges in total×103Ω×102Ω×101Ω×100Ω×10-1Ω×10-2Ω×10-3Ω×10-4Ω×10-5Ω×10-6Ω. Each range has 20 resistance values.×10-2Ω×10-3Ω×10-4Ω×10-5Ω×10-6The Ω range is a dedicated DC analog resistance box that uses a resistance network circuit to provide (0.01m Ω~ 1 μ Ω) DC analog equivalent resistance under the principle of DC voltage equivalence.
stem or root of plantsThe calibrator can verify the measuring rangeIn(1 μΩ~20kΩ)Accuracy ≤ 5 ×10-4DC resistance tester (digital bridge, micro ohmmeter).
2、 Technical indicators
- Technical indicators:≤1×10-4~1×10-2
-
Verification temperature:20℃±1℃, relative humidity: 40%~60%
Operating temperature:20℃±5℃, relative humidity: 25%~75% - chi Inch: 490×300×220
-
Basic parameters: The basic parameters of each range of the calibrator are shown in the following table:
measure Cheng |
Adjustment range |
Step disk(Ω) |
accuracy class |
Working power supply |
||
reference value |
Nominal value |
limit value |
||||
1µΩ |
0~20) µΩ |
1µΩ |
1 |
15A |
20A |
25A |
10µΩ |
0~200) µΩ |
10µΩ |
0.1 |
13A |
15A |
17A |
100µΩ |
0~2) mΩ |
100µΩ |
0.1 |
7A |
10A |
14A |
1mΩ |
0~20) mΩ |
1mΩ |
0.02 |
3A |
7A |
10A |
10mΩ |
0~200) mΩ |
10mΩ |
0.01 |
2A |
4.5A |
5A |
100mΩ |
0~2)Ω |
100mΩ |
0.01 |
700mA |
1A |
1.4A |
1Ω |
0~20)Ω |
1Ω |
0.01 |
200mA |
200mA |
450mA |
10Ω |
0~200)Ω |
10Ω |
0.01 |
70mA |
100mA |
140mA |
100Ω |
0~2) kΩ |
100Ω |
0.01 |
20mA |
20mA |
45mA |
1kΩ |
0~20) kΩ |
1KΩ |
0.01 |
7mA |
2mA |
14mA |
Power consumption: Low range resistor disk ×10-2Ω×10-3Ω×10-4Ω×10-5Ω×10-6Ω is composed of a resistor network. The power of each disc should be different.
Low range resistor disk10-2Ω, rated current ≤ 2A
Low range resistor disk10-3Ω, rated current ≤ 7A
Low range resistor disk10-4Ω, rated current ≤ 10A
Low range resistor disk10-5Ω, rated current ≤ 13A
Low range resistor disk10-6Ω, rated current ≤ 15A
Note:Please do not exceed the maximum current during use, and only for a short period of time, otherwise it may burn out the resistor.
6. Insulation resistance: The insulation resistance between the metal casing and the terminal post of the instrument under nominal conditions shall not be less than500MΩ, test voltage 500V.
7Voltage withstand test:The instrument should withstand the pressure between the metal casing and the terminal under nominal conditions50HzThe sine wave AC voltage is1000VThe test lasted for 1 minute without breakdown or discharge phenomenon.