Glove Integrity Tester GIT-MINI
GIT-MINI Glove Integrity Tester is a specialized instrument used to test the integrity of isolators/RABS system sleeves/gloves or integrated gloves. T
Product details
Glove Integrity Tester GIT-MINI

GIT-MINI Glove Integrity Tester is a specialized instrument used to test the integrity of isolators/RABS system sleeves/gloves or integrated gloves.

The GIT-MINI glove integrity tester's testing port is compatible with the isolator's operating port, and the port also integrates a differential pressure gauge for numerical display.

The GIT-MINI glove integrity tester can operate in positive pressure mode. The test results are determined by the pressure drop value per unit time. The operator determines the integrity of the testing gloves based on the reference values provided by the manufacturer.

Performance characteristics

1. Exquisite structure, portable design, no need to disassemble sleeves/gloves, convenient to use;

2. Battery powered design, no need for external power supply, easy to clean, suitable for use in clean areas;

3. Display differential pressure gauge for easy observation and operation;

4. The test gas is external compressed air and nitrogen;

5. The manual operation of the detection process is completed;

6. Can be used for leak testing of sleeve gloves before or after use of isolators;

Developed according to the ISO14644-7 standard

technical parameter

Input pressure: 0.2MPa~0.6Mpa

Inflation pressure: 500~1000Pa

Testing time: 5-8 minutes

Pressure display resolution: 1Pa

Detection accuracy (minimum aperture): 500μm

Instrument weight: 2.5kg

Host dimensions: 265 * 265 * 125mm

Power supply battery: 9V (rechargeable)

Test port specification: 255mm (other specifications can be customized)

application area

Used to test the integrity of isolators/RABS system sleeves/gloves or integrated gloves