1、 Measurement principle
Selective response of hydrogen ion concentration changes in aqueous solution using glass electrodes, andAg/AgCl/ClReference electrodes are combined together in solution to form“electrochemical cell”By measuring the changes in its electromotive force, determine the properties of the solutionpHValue.SensorLytThe electrode adopts a gel or polymer electrolyte, and the concentration of the electrolyte itself changes very little, providing a continuous and stable reference potential.
2、 Typical applications
1、 Sensolyt 700 IQ: Ordinary water/sewage
2、 Sensolyt 700 IQ SWHigh salinity wastewater such as seawater
3、 ECAElectrode: mildly polluted wastewater
4、 SEAElectrode: The electrolyte is a solid polymer that is not affected byAgClInterference, strong anti pollution ability, suitable for severely polluted wastewater, emulsifiers and suspensions, protein and sulfide containing wastewater
5、 SEA-HPElectrode: Installation of pipelines with higher temperature and pressure
6、 DWAElectrode: drinking water
7、 PtAElectrode:ORPElectrodes, municipal and industrial wastewater, emulsifiers and suspensions, wastewater containing proteins and sulfides
8、 CHEMtrac 830 MHigh temperature and high pressure wastewater
9、 PLSeries electrodes: High temperature and high pressure wastewater
3、 Advantages and Characteristics
1、 Equipped with pre signal amplification function
2、 pHandORPThe electrodes share one electrode protective shell
3、 Digital signal electrodes can store calibration data(SensorLyt)
4、 The electrode can be replaced separately, with simple operation and no need to replace the electrolyte
5、 Equipped with automatic electrode detection function: it can determine whether the glass electrode is damaged and trigger an alarm for electrode damage
4、 Technical indicators
Note:SWRefers to seawater application sensors, suitable for chloride ion concentrations greater than500mg/LHigh salinity water bodies;Ffinger181Series supporting sensors.
|
sensor |
Sensolyt 700 IQ (F*,SW*) |
CHEMtrac 830 M |
|
signal output |
digital |
Analog (can be connected with a converter)28Xgive2020System) |
|
range |
0~14 pH ±2000 mV ORP |
0~14 pH ±2000 mV ORP |
|
resolution |
0.01 pH,1 mV |
— |
|
temperature compensation |
built-inNTC,-5 ~ +60℃ |
Pt 1000perhapsPt 100 |
|
operation temperature |
0~+60℃ |
0~130℃ |
|
pressure resistance |
6~10Ba (depending on the electrode), 1 Ba(F) |
12hope earnestly |
|
protection grade |
IP68 |
— |
|
Sensor material |
stainless steel316titanium titanium+POM(SW) |
stainless steel316L |
|
power |
0.2W |
— |
|
accuracy |
±0.2 pH,±20 mV |
— |