Gas bath constant temperature oscillator, gas bath constant temperature shakerTechnical indicators:
|
model Indicator |
CHA-S |
speed range |
Start up~300rpm |
Temperature range |
Room temperature~50 ℃ |
Temp Accuracy |
±1℃ |
Amplitude |
20mm |
Oscillation |
Reciprocating |
Power |
400W |
Working Size |
520×420×200 |
time range |
0-120 minutes (or normally open) |
Power Supply |
AC 220V 50Hz |
Gas bath constant temperature oscillator, gas bath constant temperature shakerinstructions:
Using at medium speed within the speed range can extend the lifespan of the instrument.
2. The instrument should be placed on a sturdy workbench, with a clean and tidy environment, ventilation and dryness, and convenient water supply and drainage.
Before using the instrument, place the speed control knob in the minimum position and turn off the "oscillation switch".
4. Filling of culture test bottles: In order to ensure good balance performance of the instrument during operation and avoid significant vibration, all test bottle positions should be filled when bottling, and the culture solution in each bottle should be roughly equal. If there are not enough culture bottles, the test bottles can be symmetrically placed or filled with other equal amounts of solution to fill the empty spaces.
5. Connect the external power supply, place the power switch in the "on" position, and the indicator light will turn on.
6. Choose a constant temperature:
① Turn the temperature "setting temperature" selection switch to the "setting" position, adjust the temperature control knob, and the desired set temperature will be displayed digitally.
② Turn the temperature "setting temperature measurement" selection switch to the "temperature measurement" position, and the actual temperature inside the working chamber will be displayed digitally. (The red indicator light is on, indicating that the heater is working)
7. Select the timer and adjust the timer knob to the "timer" or "normally open" position.
8. Turn on the "oscillation switch" and the indicator light will turn on. Slowly adjust the speed control knob to the desired speed. When different oscillation modes need to be selected, simply change the position of the reversing switch
Before each shutdown, all switches should be placed in a non working state, the timer should be set to "zero", and the power should be cut off.
Maintenance:
1. Proper use and maintenance of the instrument can prolong its service life by keeping it in good working condition.
During continuous operation, the instrument should be regularly inspected every three months to check for any water droplets, dirt, or other debris falling onto the motor and control components; Clean the dust on the axial flow fan; Check fuses, control components, and fastening screws.